UV PROCEDURE FOR SETTING THE MONITORING STATION
1.
Gradually start the normal flow rate.
2.
Bring the unit up to normal operating pressure and temperature.
3.
Turn on the UV lamps –
wait a minimum of 15 minutes
.
4.
First set the “SET 100% LEVEL”.
See below for instructions.
5.
Then set the “Low UV ALARM LEVEL”.
See below for instructions.
Setting the SET 100% LEVEL
The
SET 100% LEVEL
potentiometer (4) is protected with a plastic screw and
O-ring set.
1) First, set the
UV DISPLAY
UNITS
button (3) to show the irradiance value
in % units.
2) Unscrew and remove the plastic screw and O-ring set.
3) Use a 2 mm (5/64”) flat blade jewelers’ screwdriver adjusting the
SET
100% LEVEL
potentiometer to 100%. This control is to be readjusted
each time the UV lamps are replaced and after first 100 hours of
operation.
4) Replace and tighten the plastic screw and O-ring set into place after
setting the SET 100% LEVEL.
Failure to do so may cause damage to the
monitoring system.
Setting Low UV ALARM LEVEL:
The
SET ALARM LEVEL
potentiometer (6) is protected with a plastic screw
and O-ring set.
1) First, set 100% level as described above.
2) Unscrew and remove the plastic screw and O-ring set.
3) Press and hold
UV ALARM SET
button (5) and use a 2mm (5/64”) flat
blade jewelers’ screwdriver to adjust the
SET ALARM LEVEL
potentiometer in %. Set the alarm level at 80%.
4) Replace and tighten the plastic screw and O-ring set into place after
setting the
SET ALARM LEVEL
.
Failure to do so may cause damage to
the monitoring system.
NOTE:
The temperature and transmission play a major role when monitoring UV intensity output in a
reactor. Therefore the set-up procedure should be carried out when the equipment is operating in a
worse case condition. The set-up of the SET 100% LEVEL and SET ALARM LEVEL should be
completed at the lowest possible water transmission and/or temperature scenario. This will prevent
you from experiencing premature alarm status if and when the feed water parameters change.
If the alarm level needs to be set at a certain absolute irradiance value, calculate this value as a
percentage with respect to the 100% level selected. As long as the SET 100% LEVEL
potentiometer (4) and the SET ALARM LEVEL potentiometer (6) are not set, the alarm value will not
be modified.