Possible Causes/Solutions:
1.
NO POWER TO THE AQUA PHASE
•
Connect Aqua Phase to a standard 20 AMP electrical
•
Press the reset button located on GFCI on the Aqua Phase power cord
2.
WATER SUPPLY IN NOT TURNED ON
•
Be certain the fill and drain connections are secure
•
Slowly turn on the water supply
3.
KINK IN THE WATER HOSE
•
Inspect water supply hose to ensure there are no sharp bends in the hose
4.
FILL-LINE FILTER SCREEN/WASHER PLUGGED
•
Turn off water supply. Disconnect Fill hose from “Fill” line connection at the Aqua Phase.
•
Inspect fill-line filter screen/washer. If plugged, contact Aqua Phase for a replacement. You
may
temporarily replace the fill-line filter screen/washer with a standard hose washer. However,
long-
term operation without this filter screen may damage the Fill valve.
5.
DEFECTIVE HIGH-LEVEL WATER CONTROL FLOAT SWITCH
•
Unplug the Aqua Phase from the power supply
•
Call Aqua Phase factory for assistance
6.
WATER SUPPLY PRESSURE IS TOO LOW
•
Turn-off water supply
•
Disconnect water supply hose from Aqua Phase
•
Determine if the water supply is adequate. Position end of hose in large bucket over drain and
turn-on your water supply. If water fills a one-gallon pail in 20 seconds, the water supply is
adequate. Reconnect hose to Aqua Phase and turn-on water supply.
7.
FILL VALVE SOLENOID IS DAMAGED OR HAS DEBRIS IN IT
•
See “Valve Inspection Procedures” in the this Aqua Phase Operating Manual
•
Remove Fill hose from “Fill” line connection at the Aqua Phase. Check that the filter
scree/washer is in in place, intact, and not plugged.
•
If problem persists, call the Aqua Phase factory