background image

 

 

 

 4/7 

1.

 

Manufacturer information 

 

AQ M-Tech AB

 operates a policy of on-going development and reserves the right to make changes 

and improvements to any of the products described in this manual without prior notice. 

Under no circumstances shall 

AQ M-Tech AB 

be held responsible for any loss or indirect damage  

howsoever caused. The contents of this document are provided as it is. 

AQ M-Tech AB

 reserves the 

right to revise this document or withdraw it at any time without prior notice. 

CE Declaration of Conformity 

Manufacturer: 

AQ M-Tech AB

 Sweden declares, that the product:  

Heater Jacket marked with CE-label conforms to the following standards:  

EN 61000-6-2:2001, EN 61000-6-4:2001, EN55011 (Group 1, Class B). 

 

Limited Warranty 

AQ M-Tech AB

 gives its end customers a warranty that the Heater Jacket is free 

from defects in its materials and operation for a period of one year from the date of purchase. 

 

Should the Heater Jacket show signs of manufacturing defects or material defects  

during the warranty period, 

AQ M-Tech AB

 will repair or replace the defective product at no charge.  

 

The warranty becomes void immediately should the Heater Jacket, during the warranty 

period, in any way be, modified, used incorrectly or be subject to tampering, abnormal working 

conditions. Such as overheating or used in any other way than what is described in this manual. 

 

The repair or replacement of other equipment in addition to the Heater Jacket can 

not be provided under the terms of this warranty. 

 

The purchaser pays the carriage charges to 

AQ M-Tech AB

AQ M-Tech AB

 can in no way be 

held liable by the purchaser for damage caused to any other equipment. 

 

 

Warning 

The Heating Jacket may only be connected to one Heater Jacket Control Unit intended for this purpose. 

Touching the Heater Jacket while in operation should only be done while wearing protective gloves 

intended for handling hot products. 

Flammable or heat sensitive materials must not be found closer than 15 cm from Heater Jacket. 

The symbol on the Heater Jacket 

“Caution!, hot surface” figure 1, means that the surface can be scorching. 

In order to avoid burn injuries, switch off the Heater Jacket and wait about 30 minutes before touching it. 

When extending the Heater Jacket cables only cables supplied by 

AQ M-Tech AB

 may be used. 

Cleaning may only be carried out when the Heater Jacket is at room temperature.  

The Heater Jacket is manufactured with different supply voltages. The supply voltage to the Heater Jacket- 

Control Unit is governed by the prescribed supply voltage for Heater Jacket. When the supply voltage to 

the Heater Jacket Control Unit is changed, Heater Jacket must also be changed.  

Heater Jacket is not approved for use in Ex-classified environments. 

 

 

 
 
 
 

Manufacturer information 

Manufacturer: 

AQ M-Tech AB 

Adress: 

Bolandsgatan 10 

 

SE-753 23 Uppsala 

 

Sweden 

Phone: 

+46 (0)18-470 29 00

 

Fax: 

+46 (0)18-10 50 04 

Web: 

www.aqmtech.se

 

E-mail:

 

 

[email protected]

 

Figure 1

 

Содержание HJSA

Страница 1: ...1 7 Heater Jacket Model HJSA HJSB Manual AQ M Tech AB ...

Страница 2: ...2 7 Heater Jacket Model HJSA HJSB Manual version 3 3 February 2017 AQ M Tech AB ...

Страница 3: ...arranty 4 Warning 4 Manufacturer information 4 2 Introduction 5 3 Installing Heater Jacket 5 4 Cleaning 5 5 Connections 5 6 Technical specification 5 7 Material selection 5 8 Dismantling the Heater Jacket 5 9 Dimensions and weight 6 10 Accessories Spare parts 6 11 Fault tracing 7 12 Dealer 7 ...

Страница 4: ...pair or replacement of other equipment in addition to the Heater Jacket can not be provided under the terms of this warranty The purchaser pays the carriage charges to AQ M Tech AB AQ M Tech AB can in no way be held liable by the purchaser for damage caused to any other equipment Warning The Heating Jacket may only be connected to one Heater Jacket Control Unit intended for this purpose Touching t...

Страница 5: ...eating cut out HJ1 155 C 155 C Maximum application temperature 140 C 140 C Maximum operating time at 140 C 400h 400h Temperature range SV 2 30 C 140 C 30 C 140 C Normal working temperature 30 C 100 C 30 C 100 C Temperature sensor type Pt 100 Pt 100 Protection class IP 65 IP 65 Ambient humidity 10 90 10 90 Steel material Stainless acid proof 316L Stainless acid proof 316L Plastic material Polyamide...

Страница 6: ... Millipore 10 138 302 104 1 6 400 115 HJSA 104 155 Millipore 5 138 179 104 1 4 200 230 HJSB 104 155 Millipore 5 138 179 104 1 4 200 115 HJSA 101 302 Sartorius Std 10 136 302 102 1 6 400 230 HJSA 101 162 Sartorius Std 5 136 162 102 1 4 200 230 HJSA 076 136 Sartorius mini 5 110 136 76 1 0 150 230 Series 3000 10 Accessories Spare parts Heater Jacket Control Unit Item Model Size WxLxH Item no Heater J...

Страница 7: ...lem Probable cause Corrective action 7 HJ does not become warm Power switch not pressed in on the HJCU No voltage Blown fuse Faulty cable connection HJ Short circuit in the HJ Faulty cable connection Temp HJCU or HJ faulty Press the power switch on the HJCU Check the incoming voltage Change the main fuse in the HJCU F1 Check the connection Contact the dealer Check the connection Contact the dealer...