background image

 

 

 

 6/7 

9.

 

Dimensions and weight 

Outside dimensions apply unless otherwise stated. The height is measured from the top section of the top ring 
to the lower section of the dome nut. 
The column filter is an example of a filter housing that matches with each HJ. 

 

Item no. 

Filter 

Width 

Height 

Top ring 

Weight 

Power 

Supply 

 

 

mm 

mm 

Id 

kg 

 

VAC 

 

 

HJSA 089 - 572 

Advanta 20" 

123 

572 

89 

5.0 

800 

230 

HJSB 089 - 572 

Advanta 20" 

123 

572 

89 

5.0 

800 

115 

HJSA 089 - 318 

Advanta 10" 

123 

318 

89 

3.0 

400 

230 

HJSB 089 - 318 

Advanta 10" 

123 

318 

89 

3.0 

400 

115 

HJSA 102 - 274 

PLT 10” 

 

136 

274 

102 

2.1 

370 

230 

HJSA 089 - 185 

Advanta 

5”   

123 

185 

89 

1.8 

200 

230 

HJSB 089 - 185 

Advanta 5”   

123 

185 

89 

1.8 

200 

115 

HJSA 073 - 130 

Junior 440   

107 

130 

73 

1.4 

100 

230 

HJSB 073 - 130 

Junior 440   

107 

130 

73 

1.4 

100 

115 

HJSA 104 - 302 

Millipore* 

10” 

138 

302 

104 

1,6 

400 

230 

HJSB 104 - 302 

Millipore* 

10” 

138 

302 

104 

1,6 

400 

115 

HJSA 104 - 155 

Millipore* 

5”  

138 

179 

104 

1,4 

200 

230 

HJSB 104 - 155 

Millipore* 

5”  

138 

179 

104 

1,4 

200 

115 

HJSA 101 

– 302 

Sartorius Std. 

10” 136 

302 

102 

1,6 

400 

230 

HJSA 101 

– 162 

Sartorius Std. 

5”  136 

162 

102 

1,4 

200 

230 

HJSA 076 

– 136 

Sartorius mini 5”  110 

136 

76 

1,0 

150 

230 

*

 

Series 3000

 

 

10.

 

Accessories/ Spare parts 

Heater Jacket Control Unit 

Item 

Model 

Size (WxLxH) 

Item no. 

 

Heater Jacket Control Unit 

 

230/115 VAC 

125x175x100 mm 

HJCU 

Heater Jacket Control Unit Communication 

3

 

230/115 VAC 

125x175x100 mm   

HJCU-COM 

Heater Jacket Control Unit Current Communic.

4

  230/115 VAC 

125x175x100 mm   

HJCU-COM-S 

Heater Jacket Control Unit Steel 

5

 

230/115 VAC 

135x212(224)x117mm 

HJCU-STEEL 

 

Supply and alarm connector without cable is included. 

Figure 2 
 
3. Heater Jacket Control Unit Communication 

– Communicates with RS-485 

4. Heater Jacket Control Unit Current Communication - Communicates with 4-20mA DC 
5. Heater Jacket Control Unit Steel made of steel 316L, Size In brackets includes mounting angles. See figure 2. 

Содержание HJSA

Страница 1: ...1 7 Heater Jacket Model HJSA HJSB Manual AQ M Tech AB ...

Страница 2: ...2 7 Heater Jacket Model HJSA HJSB Manual version 3 3 February 2017 AQ M Tech AB ...

Страница 3: ...arranty 4 Warning 4 Manufacturer information 4 2 Introduction 5 3 Installing Heater Jacket 5 4 Cleaning 5 5 Connections 5 6 Technical specification 5 7 Material selection 5 8 Dismantling the Heater Jacket 5 9 Dimensions and weight 6 10 Accessories Spare parts 6 11 Fault tracing 7 12 Dealer 7 ...

Страница 4: ...pair or replacement of other equipment in addition to the Heater Jacket can not be provided under the terms of this warranty The purchaser pays the carriage charges to AQ M Tech AB AQ M Tech AB can in no way be held liable by the purchaser for damage caused to any other equipment Warning The Heating Jacket may only be connected to one Heater Jacket Control Unit intended for this purpose Touching t...

Страница 5: ...eating cut out HJ1 155 C 155 C Maximum application temperature 140 C 140 C Maximum operating time at 140 C 400h 400h Temperature range SV 2 30 C 140 C 30 C 140 C Normal working temperature 30 C 100 C 30 C 100 C Temperature sensor type Pt 100 Pt 100 Protection class IP 65 IP 65 Ambient humidity 10 90 10 90 Steel material Stainless acid proof 316L Stainless acid proof 316L Plastic material Polyamide...

Страница 6: ... Millipore 10 138 302 104 1 6 400 115 HJSA 104 155 Millipore 5 138 179 104 1 4 200 230 HJSB 104 155 Millipore 5 138 179 104 1 4 200 115 HJSA 101 302 Sartorius Std 10 136 302 102 1 6 400 230 HJSA 101 162 Sartorius Std 5 136 162 102 1 4 200 230 HJSA 076 136 Sartorius mini 5 110 136 76 1 0 150 230 Series 3000 10 Accessories Spare parts Heater Jacket Control Unit Item Model Size WxLxH Item no Heater J...

Страница 7: ...lem Probable cause Corrective action 7 HJ does not become warm Power switch not pressed in on the HJCU No voltage Blown fuse Faulty cable connection HJ Short circuit in the HJ Faulty cable connection Temp HJCU or HJ faulty Press the power switch on the HJCU Check the incoming voltage Change the main fuse in the HJCU F1 Check the connection Contact the dealer Check the connection Contact the dealer...

Отзывы: