Reference Standard 

 

Standard Meter of Electrical Power and Energy 

 
 
 
 

Model 

RS 2130 
RS 2330 
RS 1130 
RS 1330 

 
 
 
 
 
 
 
 
 

User’s Guide 

 

Version: 9.1 

Содержание RS 1130

Страница 1: ...Reference Standard Standard Meter of Electrical Power and Energy Model RS 2130 RS 2330 RS 1130 RS 1330 User s Guide Version 9 1 ...

Страница 2: ...nu 14 3 2 8 Energy Menu 14 3 2 9 Frequency Mode Menu 14 3 2 10 History Mode Menu 15 3 2 11 Menu TALK ONLY 15 3 2 12 Menu TALK ONLY COM1 15 3 2 13 Output Frequency Response Menu 15 3 2 14 RS 232 Menu 16 3 2 15 Display Menu 16 3 2 16 Keyboard Menu 16 3 2 17 Impulse Menu 16 3 2 18 Impulse Outputs Fout1 Fout3 Menu 17 3 2 19 Impulse Input Fin Menu 17 3 2 20 Frequency Input Configuration Menu 17 3 2 21 ...

Страница 3: ... Impulse Quick Response mode 33 5 13 Harmonics 33 5 13 1 Harmonic Content 33 5 13 2 Harmonic Limitation 34 5 14 Vector Diagram 35 5 15 Meter Error Test and Meter Constant Test 35 5 15 1 COM and OPS connectors 37 5 15 2 Remote interface commands 37 5 16 Measured Values History System 39 5 17 Mode TALK ONLY 43 5 17 1 LONG format description 43 5 17 2 SHORT format description 43 5 17 3 Commands 44 5 ...

Страница 4: ...etween them Verify that the available mains voltage corresponds to specification of the unit The Safety Class I unit provided with disconnectible AC supply cable and plug may only be operated from a power socket with protective earth ground connection Any interruption or loosening of the protective earth ground conductor either inside or outside the unit or in an extension cable will cause a poten...

Страница 5: ... power consumption of the current and voltage circuits for contact error monitoring The Reference Standard makes harmonic analyze of the voltage and current spectrum and measures and computes these quantities shown in display of the device Quantity Description Code Unit Voltage Effective value of the alternating voltage U V Current Effective value of the alternating current I A Phase U Phase offse...

Страница 6: ...Reference Standard RS 2x30 RS 1x30 User s Guide RS UG91 EN www appliedp com 6 2 2 Front Panel and Rear Panel Front Panel of models RS 2x30 Front Panel of models RS 1x30 Rear Panel ...

Страница 7: ... input and output signals FOUT0 Impulse output FOUT0 FOUT1 Impulse output FOUT1 FOUT2 Impulse output FOUT2 FOUT3 Impulse output FOUT3 FIN Impulse input L1 L2 L3 Three independent measurement channels I Current clamps U Voltage clamps Lo Reference voltage clamp Hi Active current clamp In Current input clamp Out 120 A Max Current output clamp for 120 A range Out 5 A Max Current output clamp for 5 A ...

Страница 8: ... and Hi are in a differential connection against the GND i e the device measures the voltage difference between the clamps Lo and Hi Each of the Lo and Hi clamps can be loaded by operating voltage up to 500V against the GND clamp The positive orientation of the voltage difference is defined as Uhi ULo The positive orientation of the current is from the clamp In to the clamp Out The lowest error ca...

Страница 9: ...de For return to the basic mode press DEL ESC or the OK key This mode is designed for informative displaying of amplitudes of the harmonic parts of measured voltages and currents The bar on the right side of y axis is the amplitude of the 1st harmonic part Next harmonic parts are displayed in the right direction No of maximum frequency can be set in configuration menu The background noise if activ...

Страница 10: ...you enter a lower level of the menu Selected submenus are protected by a password what is typical for the calibration submenu command the label ends with special characters and the right part of the menu lebel is empty by choosing this menu item a specific command is executed parameter the label doesn t end with special characters and the right part of the menu label contains a value choosing this...

Страница 11: ...ved using OK key or skip out without accepting change If the parameter value is of numeric type the actual value is offered as start value of the editing mode The value is highlight what indicates the possibility of value editing the first used key is UP DOWN LEFT RIGHT or BS respectively delete the first used key is DEL or begin overwrite Key functions in editing mode are as follows DOWN skip to ...

Страница 12: ...he device switch off After power on of the device values of configuration parameters from the FLASH disk are loaded Def Config loading of configuration parameter values preset by manufacturer This can be used after extensive altering of configuration and unexpected behavior of the device 3 2 2 Measurement Menu Input Range Range menu Energy Energy menu Frequency Mode menu of output frequency settin...

Страница 13: ...u Keyboard keyboard menu contains items for the setting of short voice signal generation by each operation of keys BEEP and settings of keypad locking feature Keypad locking has these options LOCK SWITCH ON for locking the keypad right after each device switch on LOCK PASSWORD for password change and LOCK NOW Watchdog parameter activating deactivating the device watchdog If it is allowed then in c...

Страница 14: ...ting is switched off Voltage Factor parameter for input voltage ratio factor value The device indicates and uses the voltage value equal to product of the measured voltage value and the voltage factor value The voltage factor is used in case of input voltage transformer Current Factor parameter for input current ratio factor value Device indicates and uses the current value equal to product of the...

Страница 15: ...equal to the value of TIME BASE which is set in Measurement Menu INTEGR TIME item does not allow to change its value Reset resetting the History records 3 2 11 Menu TALK ONLY State enabling or disabling TALK ONLY Mode which serves to continuous sending of measured data to communication port COM1 to enter the settings of COM1 port in order to get more options of TALK ONLY Mode 3 2 12 Menu TALK ONLY...

Страница 16: ...if it is activated Resolution resolution setting STANDARD results with standard number of decimal digits according to the shown variable type HIGH number of decimal digits is raised by 1 comparing to option STANDARD MAXIMAL number of decimal digits is raised by 2 comparing to option STANDARD Service Info enabling or disabling ON OFF of service information Large Font parameter setting the state whe...

Страница 17: ...hannels These channels are given in Channel Lx parameters placed in the Energy menu Unit of showed value is in 1 Ws 1 Vas 1 Vars or 1 kWh 1 kVAh 1 kVArh and depends on values of Energy Factor and Factor Unit parameters Energy F Unit parameter setting the unit in which the value of parameter Energy Factor is showed Parameter is 1 Ws 1 Vas 1 Vars or 1 kWh 1 kVAh 1 kVArh 3 2 19 Impulse Input Fin Menu...

Страница 18: ... dosage is set Contact SI Mode setting the SI contact into the state CONTACT OPEN or CONTACT CLOSED or AUTO Contact SU Mode setting the SU contact into the state CONTACT OPEN or CONTACT CLOSED or AUTO Contact SI Auto entry to Contact SI Auto Mode Menu Contact SU Auto entry to Contact SU Auto Mode Menu Contact SI State setting the SI contact state open or closed Contact SU State setting the SU cont...

Страница 19: ...g DISABLED harmonic items of measured signals graph displaying if the FFT module of signal analysis is present Background parameter switching on ON or switching off OFF displaying of the noise background in the graph of harmonic items of measured signals Max freq parameter setting the maximal frequency displayed in the graph of harmonic items of measured signals 3 2 27 Contrast Menu Increase displ...

Страница 20: ...als are obligatory and small fonts at the end of words are optional command SYSTem VERSion is identical with SYST VERS but command SYSTe VERSi is not allowed letters size in the command is optional command SYST VERS is identical with sYsT VeRs 4 1 Commands Summary list of commands is available in the Appendix 4 1 1 General Commands IDN Request for the device identification string with the maximal ...

Страница 21: ...itch on of the measurement results depicting on the device display Switching off means empty display during the measurement However activated device menu is always displayed DISPlay TEXT Request for the setting of the user text which is depicted on the device display The empty string means user text extraction switch off DISPlay TEXT text User text extraction bordered by quotation marks on the dev...

Страница 22: ... no password is needed or by keyboard when password followed by key OK is typed Special password used only for unlocking keyboard can be defined in menu For unlocking keyboard can be used also calibration password needed for entering calibration submenu Function is controlled by submenu System Keyboard or by remote interface commands SYSTem KEYBoard LOCK Request for state of Keyboard Lock Answer i...

Страница 23: ...sage 2 impulse and 1 ground signals It is possible remotely via RS 232 interface or locally from the front menu of the device to detect read values of input signals modify set values on output signals set initial states values which will be set at the device start up of output signals control dosage contacts using dosage function Control connector signals can be controlled by menu item System Cont...

Страница 24: ...t Calibration menu can be saved or loaded by menu items Save Config and Def Config in Main menu or by remote interface commands SYSTem CONFig SAVe Save configuration SYSTem CONFig LOAD Load user configuration SYSTem CONFig LOAD DEFault Load default configuration Calibration settings values of parameters in Calibration menu can be saved or loaded by menu items Save Calib and Load Def Calib in passw...

Страница 25: ...y 10 minutes in first 30 minutes of device s operation is this period 5 minutes User should disable this feature always before important measuring typically before starting generation of some loadpoint and enable it again after end of this measuring after stopping loadpoint This is recommended operation of device in Meter Test Equipments and ensures autocalibration between loadpoints or in idle st...

Страница 26: ...ing of Operation Mode Replay is Operation Mode s number 0 7 5 9 2 Measuring Time Interval Time Base All measured values are evaluated in the interval of the internal Time Base The interval is defined in seconds and the default setting is 1s Time Base value is saved in configuration and recalled after switch on of device Time Base value can be controlled by menu item Measurement Time Base or by rem...

Страница 27: ...C RANGe Request for the set current measuring range CURRent AC RANGe AUTO OFF ON Switch off switch on of the automatic setting of the current measuring range CURRent AC RANGe AUTO Request for the state of the automatic setting of the current measuring range The respond is 0 if it is switched off or 1 if it is switched on CURRent AC RANGe INP Request for the state of the current input switch 40A 12...

Страница 28: ...ure IMPedance AC REAL Request for the impedance real part value MEASure IMPedance AC IMAGinar Request for the impedance imaginary part value MEASure FREQuency Request for measured frequency MEASure FREQuency CHANnel Request for channel on which frequency was measured Answer is channel number 0 5 MEASure FREQuency MODe Request for frequency measuring mode Answer is 0 automatic or 1 fixed freq MEASu...

Страница 29: ...ad using functional FUNC SHIFT key but only when this functionality is enabled in Measurement Energy RESET menu item MEASure ENERgy STARt Command for energy measurement start MEASure ENERgy STOP Command for energy measurement stop MEASure ENERgy RESet Command for energy measurement reset energy registers are set to zero MEASure ENERgy STATus Request for status of energy measurement Answer is 1 for...

Страница 30: ...quence for channels L1 L2 and L3 is ON or OFF for switched on or switched off given channel selection respectively This setting influences the summing up of the individual channels into the value of the generated frequency on the device impulse output if this is in the ENERGY mode SYSTem ENERgy IMPulse Request for the setting of the multiplication constant on the device impulse output Reply is rea...

Страница 31: ...Request for the setting of the individual channel selection for impulse output OUTx functions related to the energy Reply are three in a sequence for channels L1 L2 and L3 values 0 or 1 0 denounce for switched off and 1 for switched on given channel selection separated by a commas e g 1 0 1 SYSTem ENERgy OUTx CHANnel 0 1 OFF ON 0 1 OFF ON 0 1 OFF ON x 0 1 2 or 3 Setting of the individual channel s...

Страница 32: ... time parameter in unit seconds SYSTem ENERgy IN STATe Request for the state of the device impulse input Fin proportional to one type of energy Reply is 0 for ACTIVE POWER 1 for REACTIVE POWER or 2 for APPARENT POWER SYSTem ENERgy IN STATe 0 1 2 Setting of the state of the device impulse input Fin proportional to one type of energy Parameter is 0 for ACTIVE POWER 1 for REACTIVE POWER or 2 for APPA...

Страница 33: ...esponse mode Answer threshold 0 100 MEASure IMPulse QRESponse TREShold treshold 0 100 Setting of threshold value of Impulse Quick Response mode Parameter threshold 0 100 default value is 3 Remark Quick_Mode parameter value ON OFF is not stored into configuration and therefore it s value is remembered only until device switch off After switch on of the device has this parameter always value OFF 5 1...

Страница 34: ...udes of holded samples Reading values is not clearing or refreshing registers it means new HOLD command has to be used for new data MEASure SIGNal AMPLitude phase 0 1 2 unit 0 1 Request for amplitude of samples of defined input signal holded in memory by HOLD command Parameters phase 0 L1 1 L2 2 L3 and unit 0 U 1 I Answer samples of defined signal separated by spaces in values Amperes or Volts Eac...

Страница 35: ... has installed appropriate hardware hardware versions 1 08 and more Mode can be closed by pressing ESC or EXP buttons The basic activities of the device are not suspended device is measuring in background OPS Measurement Mode window Meter Error Test Description of OPS Measurement Mode window Left part of window 1st 3rd Line measured values measured values of active reactive or apparent power volta...

Страница 36: ... are set to correct values Actual state is indicated in end of line ON or OFF 2 or F2 Calib key for entering or leaving BARGRAPH CALIB mode of the OPS unit Actual state is indicated in end of line ON SET or OFF 3 or F3 Input key for setting input mode LED DISK or S0 4 or F4 Res key for setting resolution LOW results with two decimal places and 10000 nominal impulses 100 ppm resolution or HIGH resu...

Страница 37: ...de SYSTem OPS ERRor Request for error of measured electricity meter in OPS Measurement Mode Error is returned in a form of a real number e g 1 45 If no error is available the text string Not available is returned SYSTem OPS STATus Request for status of Local Evaluation Unit OPS Return values status codes 0 READY mode 5 MEASURE mode new error value available 1 BARGRAPH mode 6 MANUAL mode 2 CALIB mo...

Страница 38: ...standard 2 decimal places or 1 if high 3 decimal places SYSTem OPS DIVider Request for actual input divider value in Local Evaluation Unit OPS Return value is number from 1 to 65535 SYSTem OPS DIVider divider Command for setting required input divider value in Local Evaluation Unit OPS Parameter divider is number from 1 to 65535 SYSTem OPS WINDow Request for actual measurement window in Local Eval...

Страница 39: ...ected quantity Samples are stored into historical memory Value 0 means integration time equal to parameter Time Base configurable from Measurement submenu RESET reset of Measured Values History system clearing of all measured values from memory Local Configuration Configuration parameters and functions are accessible locally from this submenu SPECIAL FUNC HISTORY Remote Configuration Configuration...

Страница 40: ...s and third parameter is time string in YYYY MM DD hh mm ss xx format Example SYST HIST GET 0 1 2009 09 20 17 23 19 04730 List of quantities codes CODE QUANTITY CODE QUANTITY CODE QUANTITY 0 Active Power 1 Voltage 2 Current 3 Reactive Power 4 Apparent Power 5 Phase UU 6 Phase UI 7 Frequency 8 Distortion of Voltage 9 Distortion of Current 10 Active Energy 11 Reactive Energy 12 Apparent Energy Descr...

Страница 41: ... of all measured values of channel x 1 2 or 3 STDEV x standard deviation of all measured values of channel x 1 2 or 3 Value is in parentheses when not actual new value was measured after last calculation of STDEV value 8th Line keys 1 or F1 key for starting stopping of logging All measured values and statistics are available after stopping of logging until next start or until performing reset whic...

Страница 42: ...ing end of list during browsing or paging data 8th Line keys and keys for browsing data by one line and keys for paging data by 6 lines 3 or F3 key for selecting quantity for which measured values are shown 4 or F4 key for selecting channel 1 L1 2 L2 3 L3 for which measured values are shown Selected channel is shown in last line between 4 and 5 keys 5 or F5 key for exiting HISTORY VALUES window an...

Страница 43: ...trolled in Measure Talk Only menu or by commands 5 17 1 LONG format description codeLF channel value SI cr If Explanation of expressions codeLF measuring function long format code channel label of the measuring channel L1 L2 or L3 value real number of the measured value in the SI units in the exponential format SI label of an SI unit of the measuring function cr If End of line characters Example V...

Страница 44: ...e Quantities parameter qParam is decimal number containing binary information about enabled disabled quantities For format of parameter see detailed description in Mode TALK ONLY chapter SYSTem TALKonly COM1 QUANtities Request for setting of quantities sent on COM1 output in TALK ONLY mode Reply is decimal number containing binary information about enabled disabled quantities For format of paramet...

Страница 45: ...ase IMP AC PHAS Ph deg Q09 Impedance Real Part IMP AC REAL R ohm Q10 Impedance Imaginar Part IMP AC IMAG X ohm Q11 Frequency FREQ F Hz Q12 Distortion of Voltage VOLT AC DIST dU Q13 Distortion of Current CURR AC DIST dI Q14 Phase UU PHAS UU phU deg Q15 Phase UI in Lx channel PHAS UI phI deg Q16 Phase to phase Voltage VOLT AC PP Upp V Q17 Sum of Active Power POW AC SUM sP W Q18 Sum of Reactive Power...

Страница 46: ...ere they are opened at start and closed at the end depending on the defined initial and end status of the contacts 5 18 1 Dosage contacts SU SI SU and SI are potential free contacts located on device in Control connector During energy and time dosage operations are these contacts used to control a power source or to control the switching of current and voltage Loading capacity of the contacts max ...

Страница 47: ...initial reading of the energy registers After starting dosage with menu item or command Start the dosage is immediately started and the relay contacts are closed regarding the definition The Device now counts the predefined amount of energy or time If the amount of energy or time is reached or if the dosage is stopped earlier with menu item or command Stop the contacts are opened again to switch o...

Страница 48: ...in Ws or VAs for energy dosage SYSTem ENERgy DOSage ENERgy Request for energy in Ws or VAs for energy dosage SYSTem ENERgy DOSage ENERgy KWH energy Setting of energy in kWh or kVAh for energy dosage SYSTem ENERgy DOSage ENERgy KWH Request for energy in kWh or kVAh for energy dosage Commands for configuring SU SI contacts mode SYSTem ENERgy DOSage CONTact VOLTage MODe mode Setting of SU contact mod...

Страница 49: ... 0 contact OPEN 1 contact CLOSED or 2 contact state UNCHANGED SYSTem ENERgy DOSage CONTact CURRent AUTo STOP STATe Request for setting of SI contact state in AUTO mode before start and after end of dosage function Reply is 0 contact OPEN 1 contact CLOSED or 2 contact state UNCHANGED SYSTem ENERgy DOSage CONTact VOLTage AUTo STARt DELay delay Setting of SU contact switch delay in AUTO mode after st...

Страница 50: ...0 open 1 closed of SI contact Commands for reading status of dosage function SYSTem ENERgy DOSage Request for general status of dosage function If dosage function is active reply is percentage done energy or time and full energy or time SYSTem ENERgy DOSage STATus Request for status of dosage function Reply is 0 dosage not active 1 dosage initialized 2 dosage started and running or 3 dosage stoppe...

Страница 51: ...Reference Standard RS 2x30 RS 1x30 User s Guide www appliedp com RS UG91 EN 51 Appendix ...

Страница 52: ...C FREQuency CURRent AC CURRent AC DISTortion CURRent AC PHASe CURRent AC SUM SQuare POWer AC ACTive POWer AC REACtive POWer AC APParent POWer AC FACTor POWer AC SUM ACT POWer AC SUM REACtive POWer AC SUM APParent IMPedance AC MODule IMPedance AC PHASe IMPedance AC REAL IMPedance AC IMAGinar FREQuency FREQuency CHANnel FREQuency MODe FREQuency MODe mode FREQuency MODe VALue FREQuency MODe VALue mod...

Страница 53: ...al SAMPle OPERatingmode OPERatingmode mode HARMonicmeasmode HARMonicmeasmode mode 0 2 HARMonicmeasmode LIMit HARMonicmeasmode LIMit X 2 63 TIMebase TIMebase time SENSe VOLTage AC RANGe VOLTage AC RANGe range VOLTage AC RANGe AUTo VOLTage AC RANGe AUTo OFF ON CURRent AC RANGe CURRent AC RANGe range CURRent AC RANGe AUTo CURRent AC RANGe AUTo OFF ON SYSTem EXT RATio chnl 1 6 EXT RATio chnl 1 6 facto...

Страница 54: ...el ENERgy OUT0 CHANnel OFF ON OFF ON OFF ON ENERgy OUT0 FIXed ENERgy OUT0 FIXed freq ENERgy OUT0 FACTor ENERgy OUT0 FACTor factor ENERgy OUT1 STATe ENERgy OUT1 STATe 0 1 2 3 4 OFF ACTIVE REACTIVE FIXED APPARENT ENERgy OUT1 CHANnel ENERgy OUT1 CHANnel OFF ON OFF ON OFF ON ENERgy OUT1 FIXed ENERgy OUT1 FIXed freq ENERgy OUT1 FACTor ENERgy OUT1 FACTor factor ENERgy OUT2 STATe ENERgy OUT2 STATe 0 1 2 ...

Страница 55: ...ant ENERgy DOSage see chapter Dosage ENONly OFF ON ENONly HISTory STATe 0 1 HISTory STATe HISTory INTegrtime time HISTory INTegrtime HISTory RESet HISTory GET quantityCode chnl timeString TALKonly TALKonly OFF COM TALKonly COM1 TALKonly COM1 param TALKonly COM1 FORMat TALKonly COM1 FORMat 0 1 TALKonly COM1 QUANtities TALKonly COM1 QUANtities qParam CALib SELF SHORT FULL CALib AUTo OFF ON CALib AUT...

Страница 56: ...PS MEASure MAXCount OPS MEASure MAXCount samples OPS MEASure COUNt OPS MEASure RESult AVG OPS MEASure RESult STD OPS MEASure RESult SAMPle index KEYBoard KEYBoard BEEP KEYBoard BEEP state KEYBoard LOCK KEYBoard LOCK KEYBoard UNLock KEYBoard LCKatswitchon KEYBoard LCKatswitchon state CONTrol INPort CONTrol OUTPort CONTrol OUTPort param CONTrol OUTPort INIT CONTrol OUTPort INIT param SETup COM1 SETu...

Страница 57: ... Power Active Voltage Current Power Reactive Power Apparent Power Factor Frequency Distortion U Distortion I Phase U Phase I Sum P Active Sum P Reactive Sum P Apparent Sum U Square Sum I Square Energy Active Energy Reactive Energy Apparent Net Power Factor Impedance Module Impedance Phase Impedance Real Talk Only Fout Response Keyboard Test More Calib Per Autocalib Autocalib State COM1 Quick Mode ...

Страница 58: ......

Страница 59: ...S 1130E RS 1130S RS 1330A RS 1330E RS 1330S Product Option s all options of the above product Conforms with the following European Directives Low Voltage Directive 2006 95 EC dated 12 12 2006 EMC Directive 2004 108 EC dated 15 12 2004 Conforms with the following product standards Safety EN 61010 1 dated 2001 2002 2003 EMC EN 61326 dated 1997 Product is safe under conditions of standard application...