5.18.6 Controlling and configuring of dosage function
Dosage function can be controlled and configured by submenu “Measurement / Energy / Dosage” or
by remote commands:
Commands for controlling of dosage function:
SYSTem:ENERgy:DOSage:INIT
Initialization of dosage function.
SYSTem:ENERgy:DOSage:STARt
Start (or initialization) of dosage function.
SYSTem:ENERgy:DOSage:STOP
Stop (or cancel) of dosage function.
SYSTem:ENERgy:DOSage:CANCel
Cancel of dosage function.
Commands for configuring dosage:
SYSTem:ENERgy:DOSage:MODe {<mode>}
Setting of Dosage Mode. Parameter is number (0-
5) or code (TIME, ACTIVE …) of Dosage Mode.
SYSTem:ENERgy:DOSage:MODe?
Request for state of Dosage Mode.
SYSTem:ENERgy:DOSage:TIMe {<time>}
Setting of time (in seconds) for time dosage.
SYSTem:ENERgy:DOSage:TIMe?
Request for time for time dosage.
SYSTem:ENERgy:DOSage:ENERgy {<energy>}
Setting of energy (in Ws or VAs) for energy dosage.
SYSTem:ENERgy:DOSage:ENERgy?
Request for energy (in Ws or VAs) for energy dosage.
SYSTem:ENERgy:DOSage:ENERgy:KWH {<energy>}
Setting of energy (in kWh or kVAh) for energy dosage.
SYSTem:ENERgy:DOSage:ENERgy:KWH?
Request for energy (in kWh or kVAh) for energy dosage.
Commands for configuring SU/SI contacts mode:
SYSTem:ENERgy:DOSage:CONTact:VOLTage:MODe {<mode>}
Setting of SU contact mode. Parameter <mode> is 0(contact always OPEN), 1(contact always
CLOSED) or 2(contact AUTOmatically controlled by device).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:MODe?
Request for setting of SU contact mode. Reply is 0(contact always OPEN), 1(contact always
CLOSED) or 2(contact AUTOmatically controlled by device depending on configuration).
SYSTem:ENERgy:DOSage:CONTact:CURRent:MODe {<mode>}
Setting of SI contact mode. Parameter <mode> is 0(contact always OPEN), 1(contact always
CLOSED) or 2(contact AUTOmatically controlled by device).
SYSTem:ENERgy:DOSage:CONTact:CURRent:MODe?
Request for setting of SI contact mode. Reply is 0(contact always OPEN), 1(contact always CLOSED)
or 2(contact AUTOmatically controlled by device depending on configuration).