57
a method that ignores fill level, electrolysis, electrode characteristics, etc.,
and features a microprocessor to perform temperature compensation. In
simpler instruments, conductivity values are usually assigned an average
correction similar to that of KCl solutions for correction to 25°C. The
correction to an equivalent KCl solution is a standard set by chemists
that standardizes the measurements and allows calibration with precise
KCl solutions. In the Ultrameter
III
, this correction can be set to other
solutions or tailored for special measurements or applications.
B. Tempco Variation
Most conductivity instruments use an approximation of the temperature
characteristics of solutions, perhaps even assuming a constant value.
The value for KCl is often quoted simply as 2%/°C. In fact, KCl tempco
varies with concentration and temperature in a non-linear fashion. Other
solutions have more variation still. The Ultrameter
III
uses corrections
that change with concentration and temperature instead of single
average values. See Chart 1.
C.
An Example of 2 different solution selections and the
resulting compensation
How much error results from treating natural water as if it were KCl at
15°C?
A tap water solution should be compensated as 442 with a tempco of
1.68 %/°C, where the KCl value used would be 1.90 %/°C.
Distributed by Applied Membranes, Inc. www.appliedmembranes.com
Содержание MYRON L COMPANY Ultrameter III 9PTK
Страница 2: ...Distributed by Applied Membranes Inc www appliedmembranes com ...
Страница 4: ...ii Distributed by Applied Membranes Inc www appliedmembranes com ...
Страница 73: ...Distributed by Applied Membranes Inc www appliedmembranes com ...
Страница 75: ...Distributed by Applied Membranes Inc www appliedmembranes com ...