User mode calibratiOn linc™ FUnctiOn
The Linc™ function allows easy calibration when in
User mode and the
user does not have a user standard solution to calibrate the instrument.
This function will ensure more repeatable and accurate measurements
than many other calibration methods. It is recommended that this
function be used to provide the highest degree of confidence when
the Ultrameter
III
is used in User mode. When Linc is used, the User
mode is linked to another standard, i.e., if User and KCl are linked,
a KCI standard solution is used to calibrate the instrument. It is that
simple.
A. Calibration of Ultrameter
III
for use in User mode
1.
Press
or
key.
2.
Calibrate the unit using a Standard Solution (ref.
CALIBRATION, pg. 26).
3.
Place the Ultrameter
III
in User mode (ref. SOLUTION
SELECTION, pg. 22).
4.
Verify/Set the calibration linc. (See below – Setting “User”
Mode Calibration “Linc”.)
B.
Setting User mode Calibration “Linc”
The Linc function sets or “links” the calibration gain factor of a Standard
Solution to the User solution mode. Once set, the “
linc
” will stay intact
with future calibrations unless the Linc has been canceled. For more
information on canceling the User mode Calibration Linc refer to the
section “Canceling User mode Calibration “
linc
””, pg. 46.
Distributed by Applied Membranes, Inc. www.appliedmembranes.com
Содержание MYRON L COMPANY Ultrameter III 9PTK
Страница 2: ...Distributed by Applied Membranes Inc www appliedmembranes com ...
Страница 4: ...ii Distributed by Applied Membranes Inc www appliedmembranes com ...
Страница 73: ...Distributed by Applied Membranes Inc www appliedmembranes com ...
Страница 75: ...Distributed by Applied Membranes Inc www appliedmembranes com ...