If none of the Solution Selection icons are displayed (i.e., KCl,
NaCl or 442), nothing has been linked to User mode.
6.
Press
key to accept the setting. Pressing any of the
measurement keys will exit without changing the setting. User
mode “
linc
” is now complete. The User mode will now use
the calibration gain constant used for the calibration of the
Standard Solution as outlined above.
C.
Canceling User mode Calibration “Linc”
The Ultrameter
III
must be in User linked mode in order to cancel the
“
linc
” (ref. Solution Selection, pg. 22).
1.
Press “Linked” measurement key
,
or
.
Two solution icons will be shown in the left side of display —
“
User
” and another, e.g., “
Kcl
”.
2.
Press
key until the menu “
linc
” appears (see Figure 27).
3.
Press
key; the instrument will display both “
sel
” and
the “
User
” Icon.
4.
Press
key until “
User
” is the only solution icon being
displayed.
5.
Press
key.
6.
The User mode calibration “
linc
” has now been canceled.
nOtes:
1.
To maintain repeatability, use the same standard solutions for
future calibrations.