BigDye
™
Kit Standard Run Method
Cycle/Step
Stage
Reaction
volume
(µL)
Cover
temp. (°C)
1
2
3
4
Cycles
1
25
1
—
50
105.0
Step 1
RampRate
6.0
6.0
6.0
—
Temperature
96.0
96.0
4.0
Hold Time
60
10
Infinite
Step 2
RampRate
—
6.0
—
—
Temperature
50.0
Hold Time
5
Step 3
RampRate
—
6.0
—
—
Temperature
60.0
Hold Time
240
Elongase
®
2-Step Method
Cycle/Step
Stage
Reaction
volume
(µL)
Cover
temp. (°C)
1
2
3
4
Cycles
1
30
1
—
50
105.0
Step 1
RampRate
6.0
6.0
6.0
—
Temperature
94.0
94.0
4.0
Hold Time
30
30
Infinite
Step 2
RampRate
—
6.0
—
—
Temperature
68.0
Hold Time
60
Elongase
®
3-Step Method
Cycle/Step
Stage
Reaction
volume
(µL)
Cover
temp. (°C)
1
2
3
4
Cycles
1
30
1
—
50
105.0
Step 1
RampRate
6.0
6.0
6.0
—
Temperature
94.0
94.0
4.0
Hold Time
60
30
Infinite
Appendix E
Predefined run methods
Location of power point and ports on the instrument
E
SimpliAmp
™
Thermal Cycler User Guide
75