AmpliTaq Gold
™
360 Method
Cycle/Step
Stage
Reaction
volume
(µL)
Cover
temp. (°C)
1
2
3
4
Cycles
1
35
1
—
50
105.0
Step 1
RampRate
6.0
6.0
6.0
—
Temperature
95.0
95.0
72.0
Hold Time
300
15
420
Step 2
RampRate
—
6.0
6.0
—
Temperature
55.0
4.0
Hold Time
30
Infinite
Step 3
RampRate
—
6.0
—
—
Temperature
72.0
Hold Time
60
AmpliTaq Gold
™
Fast PCR Method
Cycle/Step
Stage
Reaction
volume
(µL)
Cover
temp. (°C)
1
2
3
4
Cycles
1
35
1
—
50
105.0
Step 1
RampRate
6.0
6.0
6.0
—
Temperature
95.0
96.0
72.0
Hold Time
600
3
10
Step 2
RampRate
—
6.0
6.0
—
Temperature
62.0
4.0
Hold Time
3
Infinite
Step 3
RampRate
—
6.0
—
—
Temperature
68.0
Hold Time
5
Appendix E
Predefined run methods
Location of power point and ports on the instrument
E
SimpliAmp
™
Thermal Cycler User Guide
71