7-25
Entering an Idle State Setpoint Value
The
idle state setpoint is the temperature of the block when a
program is not running (the idling temperature). The
Configuration function lets you change the idle state setpoint
value. The default value for the idle state setpoint is 25
degrees C. The instrument will automatically heat or cool to this
temperature while idling and the Heating light on the control
panel will flash.
To edit this configuration parameter:
1
Select CONFIG on the Utilities menu as described on page 7-3.
The following display appears:
2
Press
ENTER
, then press
STEP
four times. The following display
appears:
3
To change the idle state setpoint, press the appropriate numeric
keys, then press
ENTER
. For your safety, set the idling temperature
close to room temperature.
4
Press
STEP
to display the next configuration parameter or press
BACK
to display the previous configuration parameter. Press
STOP
to return to the first Configuration display.
Configuration
EDIT-PRINT
Idle state setpoint
25C
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание GeneAmp PCR System 9600
Страница 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 79: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 251: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 275: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 281: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 294: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...