5-36
Setpoint #3 in a typical cycle program is used for
primer extension. This display sets the ramp time,
target temperature, and hold time for the third
setpoint (Setpt #3). The default values are shown in
the display to the left.
You can accept any of the default values for Setpoint
#3 by pressing the
ENTER
key, or you can change
the values using the same procedure as that for
Setpoint #1 (see above).
A display will appear for each setpoint up until the
total number of setpoints specified in the first
program display. The default temperature for
setpoints in CYCL programs with more than four
setpoints is 0.0 degrees. A CYCL program call have
up to nine setpoints.
Specifies the number of times the complete cycle
will be repeated. Type the number of cycles (1-99),
then press
ENTER
.
This display also allows you to program a pause
after a setpoint and cycle that you will specify in
subsequent displays. A pause would be required,
for example, to add a reagent. A pause may also be
used to add or remove sample tubes during a run.
Press the
OPTION
key to change the NO response
(default) to a YES and vice versa. When the desired
response is displayed, press
ENTER
.
Setpt #3 Ramp 0:00
72.0C Hold
0:30
Setpt #n Ramp 0:00
72.0C
Hold 0:30
Total cycles = 25
Pause during run? NO
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание GeneAmp PCR System 9600
Страница 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 79: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 251: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 275: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 281: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 294: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...