Description of AUTO Program Displays
Defines the number of incubation temperatures or
setpoints used in the PCR cycle. A typical PCR
cycle consists of three (3) setpoints; therefore, 3 is
the default value.
Type the number of setpoints you wish in your DNA
amplification experiment (1-9), then press
ENTER
.
This display sets the following values for the first
setpoint (Setpt #1):
♦
Setpoint temperature. The default value is 94.0
degrees C.
♦
Hold time: the length of time the samples will be
maintained at the specified temperature. The
default value is 30 seconds.
Note
If you previously used the DNA Thermal Cycler,
note that you will be using shorter hold times for the
GeneAmp PCR System 9600. Refer to page 5-2 for more
information.
You can accept any of the default values for Setpoint
#1 by pressing the
ENTER
key, or you can change
the values using the following steps.
3 Temperature PCR
Setpt #1 94.0C
Hold for 0:30
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание GeneAmp PCR System 9600
Страница 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 79: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 251: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 275: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 281: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 294: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...