![Applent Instruments Anbai AT510 Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/applent-instruments/anbai-at510/anbai-at510_user-manual_2954870018.webp)
18
AT510x DC Resistance Meter
4.
Comparator
This chapter provides the following information:
Turn ON/OFF
Setup comparator record number file
Setup nominal value and limit value
Setup Beep
4.1
Turn
ON/OFF
Comparator
AT510X’s comparatorcan be turned OFF by pressing Comp key.
When the comparator function is OFF, sorting system will no longer work and the
Handler interface signals on the comparator output will be shut down.
4.1.1
Setup
Comparator
Record
File
Number
(
invalid
at
AT510L/AT510M
version
)
1
.
Press Shift Comparator to the setup comparator.
2
.
Rec.09 indicator at the lower right corner of screen
flashes means you can setup record number.
3
.
Press < >or twist knob to choose the record
number which built-in 30 files.
4
.
Press digits to input value.
a. Press Enter to input nominal value and limit
value.
b. Press Esc to exit setup comparator and the
record file is saved.
4.1.2
Setup
nominal
value
and
limit
value
1
.
Press Comparator key to the setup comparator
status.
2
.
Repeat steps from 2-3 and press Enter to input the
comparator status and the digits are blink.
3
.
Press < > keys or twist knob to choose the
resistance’s nominal value (1st line), upper limit
value (2nd line) and lower limit (3rd line)
4
.
Press digit number keys or Enter key to input the
value.
5
.
Press Enter key or
Shift
+ Unit keys to input the
current value.
6
.
Repeat steps from 3~5 to input the other values.
7
.
PressEsckey to exit the setup and return the
measurement status.
TIP:
Using skill
:
The present value is flashed ,you can press digits to input the value
directly and it’s not necessary to press the Enter key first to enter the input status.
NOTE
:
In the number input status, press Enter key, the unit is 1.
For example
:
10+Enter :input value is :10
。
In the range hold status (“AUTO” indication is OFF), Meter will select the range by
the upper limit value when it is exit setup status.