Applent Instruments Anbai AT510 Скачать руководство пользователя страница 17

Start up

 

17

    Indicator lighted means the keypad is locked. 

3.4.9

 

20mV

 

limit

The

 

function

 

is

 

invalid

 

in

 

the

 

AT510x’s

 

version.

 

The function is invalid in the AT510x’s version. 

3.4.10

 

Temperature

 

Compensation

 

Function

 

AT510PRO, AT510: Standard function.AT510SE: Optional function.AT510L/510M: 
Invalid function. 
The temperature probe is plugged into the jack

the function is enabled. 

Resistance value of temperature compensation works when the meter is equipped with 
temperature probe temperature. 
When the temperature probe is properly plugged into the jack and turn on the meter, the 
temperature value will display in the 2

nd

 line. 

Press 

Shift

 Relative keys to turn ON/OFF temperature relative function. The REL 

indication is ON, the function works. 
Input temperature coefficient

α

 

1

 

Press 

Shift

 Relative keys to input password

11111

”(

5 digits

and setup the 

α

2

 

For example, copper material temperature coefficient of 20 

 is 0.00393,enter 

0.393%. 

3

 

Press 

Shift

 Relative keys to input password“22222”

digits and setup 

temperature status. 

4

 

Press number keys to input the temperature value. The normal relative temperature 
is room-temperature ( 20

º

C ). 

Compensation formula

 

 

T0 

 Reference temperature

take 20

℃)

 

 Current temperature 

α

  Temperature coefficient of reference temperature 

F1 

  Without compensation value 

F2 

 Temperature compensated value 

 

3.4.11

 

Remote

 

Control

 

Remote control function is used for opening series communication interface. 
The 

indication is ON,Remote control function is ON. 

1

 

Press Shift +Remote to the Remote control status. 

2

 

Select RS-232 and setup Bond rate..(Normal: 9600) 

3

 

Press Esc key to the measurement status. 

4

 

Press Enter key to save and exit. 

 

3.4.12

 

Adjust

 

VFD

 

Brightness

 

Press Shift Brightness key to adjust VFD Brightness. 
The first line of VFD displays “VFD-LT” and the 2

nd

 line shows current brightness level. 

Press <, > or turn the Knob to change a new level. 
Press Enter to save and exit to discharge state. Press Esc to exit to discharge state but not 
be saved. 
Brightness includes 8 levels: 
0(dark) ~ 7(bright) 

Содержание Anbai AT510

Страница 1: ...trict Changzhou Jiangsu China 213014 Tel 0519 88805550 89966117 89966227 Fax 0519 89966550 http www applent com Sales Email sales applent com Tech Email tech applent com 2005 2014 Applent Instruments Inc Rev D3 AT510PRO 510 510SE 510L 510M DC RESISTANCE METER User s Manual ...

Страница 2: ...t To avoid electric shock hazard the instrument chassis and cabinet must be connected to a safety earth ground by the supplied power cable with earth blade DO NOT Operate In An Explosive Atmosphere Do not operate the instrument in the presence of inflammable gasses or fumes Operation of any electrical instrument in such an environment constitutes a definite safety hazard Keep Away From Live Circui...

Страница 3: ...o a service facility designated by Applent The buyer shall prepay shipping charges to Applent and Applent shall pay shipping charges to return the product to the Buyer However the Buyer shall pay all shipping charges duties and taxes for products returned to Applent from another country Applent warrants that its software and firmware designated by Applent for use with an instrument will execute it...

Страница 4: ...st side 14 3 4 2 Test Current Modes 15 3 4 3 Range 15 3 4 4 Speed 16 3 4 5 Calibration Short Zeroing 16 3 4 6 Relative Value for Temperature Compensation 16 3 4 7 The function is invalid in the version 16 3 4 8 Keypad Locked 16 3 4 9 20mV limit The function is invalid in the AT510x s version 17 3 4 10 Temperature Compensation Function 17 3 4 11 Remote Control 17 3 4 12 Adjust VFD Brightness 17 4 C...

Страница 5: ...npacking and Inspection 5 6 1 High Current Test Mode 24 6 2 Low Current Test Mode AT510PRO AT510 25 6 3 Plus Current Test Mode AT510PRO AT510 25 6 4 General Specification 26 6 5 Dimensions 27 7 Models 28 ...

Страница 6: ...as not been damaged 2 Referring to Table 1 1 check that all packaged items supplied with the meter have been provided as per the specified optioned Table1 1 Included Accessories NAME QTY REMARK User s Manual 1 AC Power Cord 1 Fuse 2 0 5A Slow Blow Kelvin Test Leads 1 ATL501 Test Report 1 Product Warranty Certification 1 1 2 Power Supply Confirm that the power supplied to the AT510x meets the follo...

Страница 7: ...uirements Ensure that the operating environment meets the following requirements Temperature Range 0 C 55 C Humidity Range 95 RH 40 C 1 5 Cleaning To prevent electrical shock disconnect the AT510x power cable from the receptacle before cleaning Use a dry cloth or a cloth slightly dipped in water to clean the casing Do not attempt to clean the AT510x internally WARNING Don t Use Organic Solvents su...

Страница 8: ...er 1 6 How to Remove the Handle A handle kit is attached to the AT512 Figure 1 1 Hand Handle 1 Retracted Carrying Position 2 Extended Remov e Handle Lift the handle perpendicular to the unit while pulling it in the direction of ...

Страница 9: ... control data acquisition and analysis Computer remote control commands compatible with SCPI Standard Command for Programmable Instrument Programmable Instruments standard command set complete and efficient remote control and data acquisition functions AT510x measures of high medium and low value resistor various switch contact resistance connector contact resistance relay line package and the con...

Страница 10: ...indow size 98mm 58mm Calibration Short Zeroing Comparator sorting function Built in 30 sorting files AT510Pro AT510 AT510SE and Output 3 levels HI IN and Low or GD NG display beep sound Display Direct display on the VFD displayer Output Output the sorting results by the Handler interface or RS232C interface Beep Setup the sorting results and turn ON OFF Beep Beep and VFD Brightness can be Adjusted...

Страница 11: ...3 2 VTH 6 Table 3 1 Front panel description No Function 1 Power Switch Down ON Up OFF 2 Keypad I 3 Shift key Yellow Press the key the shift indication is ON and the keys are the 2nd Function Yellow 4 Keypad II 5 UNKNOWN Terminal 6 Arrow keys Select the options 7 Knob To choose function and input number value 3 1 2 Keypad ASSUMER On the front Panel Black Words on Button represents 1st Function Oran...

Страница 12: ...el Orange Words on Panel represents 2nd Function The keys will be 2nd function while the Shift mark on VFD is on Table 3 3 Secondary Functions description Key Function Comparator Open or close comparator function 20mV limited The function is invalid in these AT510x versions Test Current Modes Select test current modes High current H Cur Low current L Cur and Plus current P Cur Relative Value Relat...

Страница 13: ...mparator function is ON SET Setup comparator status HIGH Upper limit value LOW Lower limit value Keypad Locked DH Data Hold LEVEL Test Current Mode EX External Trigger Mode TRIG Manual Trigger Mode F M S Speed Fast Medium and Slow DELAY The function is invalid in the version Display Range number in the measurement status and display the sorting number in the comparator setup status Shift Shifted t...

Страница 14: ...not enter the measurement state 3 3 3 Power up Defaults The power on default will be the last configuration you saved 3 3 4 Warm up Time AT510x is ready to be used as soon as the power up sequence has completed However to achieve the accuracy rating warm up the instrument for 30 minutes 3 4 Measurement configuration 3 4 1 Connection Test side Red logo testing BNC cables into the H in the first rot...

Страница 15: ...e present current mode displays in the 3rd line H CUR High Current Mode L CUR Low Current Mode P CUR Plus Current Mode Detailed test current modes see Appendix A 3 4 3 Range In the automatic range status AT510x will choose the fit range by the following table Table 3 4 Range no reference Resistance and Range change process AT510Pro NO Reference Resistance Up Down 1 10m 30m 300m 3 30 300 3k 30k 300...

Страница 16: ...and make the test clips short circuit like the following way before zeroing Figure 3 7 Right Way Figure 3 8 Wrong Way 2 Press Enter key to Calibration The meter zeroing all ranges in the automatic range or the present range in the hold range Data will be saved in the nonvolatile memory when zeroing is over 3 Press Esc or when zeroing is over return to the measurement status 3 4 6 Relative Value fo...

Страница 17: ...f 20 is 0 00393 enter 0 393 3 Press Shift Relative keys to input password 22222 5 digits and setup temperature status 4 Press number keys to input the temperature value The normal relative temperature is room temperature 20ºC Compensation formula T0 Reference temperature take 20 T Current temperature α Temperature coefficient of reference temperature F1 Without compensation value F2 Temperature co...

Страница 18: ... exit setup comparator and the record file is saved 4 1 2 Setup nominal value and limit value 1 Press Comparator key to the setup comparator status 2 Repeat steps from 2 3 and press Enter to input the comparator status and the digits are blink 3 Press keys or twist knob to choose the resistance s nominal value 1st line upper limit value 2nd line and lower limit 3rd line 4 Press digit number keys o...

Страница 19: ...ress Esc key to exit the setup and back to the measurement state 4 Press Enter key to end the setup and back to the measurement status 4 1 5 Flow of comparator work When the comparator function is enabled the measure value compares with the upper limit and the lower limit values Sorting Flow Limit Lower value Rx Upper Limit Pass Display GD and IN Rx Limit Lower value Fail Display NG and LO Rx Limi...

Страница 20: ... system controller 5 1 Pin Assignment Figure 5 1 Pin Assignment Power Supply PIN Signal Description 1 NC Unused 2 EXT DC 5V External DC Voltage 5V 3 GND Power ground 4 NC Unused External control signal input 5 TRIG An external trigger signal 6 NC Unused 7 NC Unused 8 COMP 4 Comparator record files Built in 30 files 9 COMP 3 10 COMP 2 11 COMP 1 12 COMP 0 Comparator record files COMP 4 0 NO COMP 4 0...

Страница 21: ...p2 With the button pushed down insert the cables into the holes Step3 Release the button and the cables secured Recommended wire Single strand 0 65mm dia AWG 22 Multi strand 0 32 mm2 AWG 22 Usable limits Single strand 0 32 to 0 65mm dia Multi strand 0 08 to 0 32mm2 Standard insulation stripping length 10mm Button pressing tool Blade screwdriver shaft diameter 3 tip width 2 6mm To prevent damage to...

Страница 22: ...ector by using a photo coupler The voltage of each output is obtained by connecting pull up resistors inside or outside of the AT520 5 4 3 Power supply The power supply for the judgment output signal pull up and that for the operation output signal pull up and input signal drive can be set separately You can select from 3 3V to 35V external power supply Figure 5 3 Typical Circuit Diagram of Handle...

Страница 23: ...Handler Interface 23 5 5 Timing Chart TRIG EOC OUTPUT t2 t1 t3 t4 有有 t1 Measurement time Fast 66 7ms Medium 200ms Slow 330ms t2 1 s t3 200 s t4 0 s ...

Страница 24: ...30Ω 30 000Ω 1mΩ 0 1 0 05 2 6 7mA 1V 5 300Ω 300 00Ω 10mΩ 0 1 0 05 2 670uA 5V 6 3kΩ 3 000kΩ 100mΩ 0 1 0 05 2 670uA 5V 7 30kΩ 30 000kΩ 1Ω 0 1 0 05 2 67uA 5V 8 300kΩ 300 00kΩ 10Ω 0 1 0 05 2 6 7uA 5V 9 3MΩ 3 0000MΩ 100Ω 0 1 0 05 2 0 67uA 5V 10 20MΩ 20 000MΩ 1kΩ 0 1 3字 0 1 5 3V AT510 Range Maxim Display Value Resolution Fast Medium Slow Test Current Open circuit terminal voltage 1 30mΩ 30 000mΩ 1 Ω 0 1 ...

Страница 25: ...tion Fast Medium Slow Test Current Open circuit terminal voltage 1 30Ω 30 000Ω 1mΩ 0 1 0 1 2 6 7mA 1V 2 300Ω 300 00Ω 10mΩ 0 1 0 1 2 670uA 5V 3 3kΩ 3 000kΩ 100mΩ 0 1 0 1 2 670uA 5V 4 30kΩ 30 000kΩ 1Ω 0 1 0 1 2 67uA 5V 5 300kΩ 300 00kΩ 10Ω 0 1 0 1 2 6 7uA 5V 6 3MΩ 3 0000MΩ 100Ω 0 1 0 1 2 0 67uA 5V 7 20MΩ 20 000MΩ 1kΩ 0 1 3字 0 1 5 3V 6 2 Low Current Test Mode AT510PRO AT510 Only for range 300m 3k oth...

Страница 26: ...y Value 30000 Test Signal 30m 300k Constant Current 2M 20M Constant Voltage Trigger Mode Internal External and Remote Range Automatic and Range Hold Correction Short Zeroing Comparator Output NG LO GD IN NG HI built in 30 sorting files Beep GD NG Open Close Function and Adjust volume Test Terminal 4 terminals with ground shielding Interface Handler interface AT510PRO AT510 AT510SE RS232 interface ...

Страница 27: ...Specification 27 6 5 Dimensions ...

Страница 28: ... Internal Manual External Remote Internal Manual Internal Manual Internal Manual Test Current Mode High Low Plus High High Comparator Record File 30 Files 1 File 1 File Temperature Relative Function Standard Optional Interface RS232C HANDLER RS232C Optional HANDLER Optional Comparator Hi IN Lo GD NG Range Mode Automatic Range Hold Display 30000 digitals 4 2 3 VFD Display Beeper GD NG OFF adjust th...

Отзывы: