AT510x DC Resistance Meter
9
1M
10
10M
Press Range < > keys to choose the fit range.
Tip:
Range Hold is help increasing the measurement rate..
In the “automatic range” status, the meter should be calibration (zeroing) when it can’t
choose the fit Range.
Full calibration (Zeroing) in the “Calibration”.
In the range hold status (“AUTO” indication is OFF), the meter will choose the fit range by
the upper limit value.
3.4.4
Speed
Press Rate key to choose the following speed
:
Fast: Fast Sampling
Medium: Medium Sampling
Slow: Slow Sampling
Full Sampling Speed in the Appendix A.
3.4.5
Calibration
(Short
Zeroing)
.
1 Press Clear key to the Clearing status and make the test clips short-circuit like the
following way before zeroing.
Figure 3‐7
Right Way
Figure 3‐8
Wrong Way
.
2 Press Enter key to Calibration.The meter zeroing all ranges in the automatic range or
the present range in the hold range.
Data will be saved in the nonvolatile memory when zeroing is over.
.
3 Press Esc or when zeroing is over, return to the measurement status.
3.4.6
Relative
Value
for
Temperature
Compensation
AT510pro, AT510 (Standard), AT510se (Optional), AT510L, AT510M (invalid)
Press
Shift
Relative key to Turn ON/OFF the relative function. The REL indicator is OFF;
the value is the real measurement value.
3.4.7
‐
%
(The
function
is
invalid
in
the
version.)
3.4.8
Keypad
Locked
Press Shift Key Lock key to lock/unlocked the keypad.
The Key Lock key is only available and the other keys are in vain when the keypad is
locked.