5.5.2.4 CC/CV Prior
This menu provides CC/CV priority function allows the user to select suitable mode correspond
to test requirement, let the output be voltage high speed (CV Prior) or current no overshoot
(CC Prior) mode. Available setting range for delay time is from 0.001s through 10s.
5.5.2.5 Hi-Variation
There is a internal dummy load that can be enabled to increase the speed of the voltage fall
time for high speed test applications. The feature should only be used with caution, as it is not
designed for all applications. The default setting is OFF.
DO NOT enable this function for application such as battery or similar type loads.
Note
5.5.3 Load-2QO
5.5.4 Sink On/Off
Press the Disp 1 of 2 soft key to turn to the next page.
5.5.5 Waveform
The Waveform soft key displays a visual time domain representation of the output.
5.5.6 Limitation
The limitation soft key displays the user limit setting screen. User limits are useful to protect a
unit under test from operator error by setting the upper and or lower parameter value limits. To
set any of the available user limits, use the cursor or touch screen to select one of the available
limit set value.
●
Vlimit
Ilimit
●
Plimit
●
Rlimit
●
5.5.6.1 Vlimit
This menu allows the user to set the upper and or lower parameter value limits for voltage.
The user limits must fall within the hardware limits of the power source, which are: