background image

15

High pH results

Water sample temperature 

above 100 °F (38 °C) 

interferes with pH reagent

For water samples over 100 

°F (38 °C) subtract 0.3 from 

pH result or, for the most 

accurate result, wait until 

water sample is below 90 °F 

(32 °C) to test

Disk not spinning

Lid open, meter not 

powered on, low battery, 

disk or disk cover pressed 

down too tightly on hub

Close lid, power on meter, 

charge the battery or plug 

meter into a stable power 

source, remove the disk/disk 

cover and place back in the 

chamber more gently

Fast electrical transients 

may disrupt operation of the 

API AQUASPIN meter

Restart the test to resume 

normal operation

Trouble connecting to 

computer by USB

Broken connection

Press and hold power button 

for 1 second.

Meter won’t turn on

Needs reset or evaluated

Contact Support

METER CHECK DISK TROUBLESHOOTING GUIDE

Do not attempt to separate the components of the Meter Check Disk (Code AP-

1705). The Meter Check Disk consists of a disk with a permanently attached cover.
Do not 

fi

 ll the Meter Check Disk with water. Water is not used during the meter 

check procedure.

Range Check Procedure

The Meter Check Disk is used to verify the performance of a meter. It is not used 

to recalibrate the meter. Meters are calibrated at the time of manufacture. The 

meter is performing satisfactorily if readings using the Meter Check Disk are within 

the ranges provided on the Meter Check Disk pouch. Range speci

fi

 cations 

are speci

fi

 c to the disk identi

fi

 ed by the serial number on the pouch. The range 

speci

fi

 cations will vary from disk to disk. The exact readings from a speci

fi

 c disk may 

vary from meter to meter. 

1.  Follow the Cleaning procedure on page 21 to clean the light chamber and 

optic lenses. 

2. Tap   select Disk Series.
3. Tap 

MCD 

to select Meter Check Disk. 

4. Tap to return to the test screen.
5.  Remove the Meter Check Disk from the foil pouch. DO NOT remove the black 

cover from the disk.

6.  Insert the Meter Check Disk. Close the lid.
7. Tap 

 to start test.

8.  Results will be displayed.
9.  Compare the results on the display to the values on the chart shown on the 

pouch.

10. Replace the Meter Check Disk in the foil pouch for storage.

If the results are not within the range shown on the pouch, contact Support.

Содержание AP-3580

Страница 1: ...Simple Precise Digital Water Testing...

Страница 2: ...2...

Страница 3: ...k Cover 12 Meter Check Disk 12 Usb Cable 12 Battery 12 Settings 13 TROUBLESHOOTING Meter Troubleshooting Guide 14 Meter Check Disk Troubleshooting Guide 15 AQUASPIN Web Application Troubleshooting Gui...

Страница 4: ...re register and clicking Create account 1 You will need to enter the serial number of your meter to create your account The serial number can be found on the bottom of your AQUASPIN meter or on your w...

Страница 5: ...t time to doing the update and nothing else on your computer to allow the update to not be disrupted PC CONNECTION When the API AQUASPIN meter is connected to a PC via USB cable the onboard touchscree...

Страница 6: ...O NOT overfill the disk If the disk is overfilled sample water will flow out of the overflow hole in the center of the disk If this happens be sure to dry the disk completely before running the test O...

Страница 7: ...e air more than necessary Do not transport the meter with a disk in the chamber Store disks at 70 80 F 21 27 C METER When a filled disk is placed in the chamber with universal disk cover and the lid i...

Страница 8: ...election tap the icon or word on the screen with a fingertip fingernail pencil eraser or stylus Gently wipe smudges from the screen with the Cloth Wipe Do not touch the screen with a sharp object Do n...

Страница 9: ...ount enter their email address and tank type under the REGISTER NEW CONSUMER ACCOUNT section 6 Next select the disk type the disk type can be found on the foil packet Then hit RUN TEST button on the c...

Страница 10: ...s to turn the meter off NOTE Remove salt residue daily Salt will damage the meter and cause inaccurate results See the CLEANING section in the manual NOTE For water samples over 100 F 38 C subtract 0...

Страница 11: ...esh Q104 21 43 23 May Fresh Q104 20 41 22 May Fresh Q104 20 36 22 May Fresh Q104 Test History View Selected Results Delete Selected Results History Settings Select Auto Save Go to Top Go to Bottom Scr...

Страница 12: ...Check Disk is inserted in the meter and Rotary Calibration is chosen from the Settings Menu the alignment of the hub and disk is evaluated The results are analyzed and reported as Pass or Fail For us...

Страница 13: ...at any time Set Language There is one language option English Tap to exit to the Settings menu Calibration Tap to run an angle calibration to evaluate the alignment of the hub and disk Tap to exit to...

Страница 14: ...filled disk or air bubble Solution Fill disk correctly see FILLING Range Error Raw data out of range Contact Support Output Error Decreased light intensity Possibly dirty lens Clean lens see CLEANING...

Страница 15: ...heck Disk with water Water is not used during the meter check procedure Range Check Procedure The Meter Check Disk is used to verify the performance of a meter It is not used to recalibrate the meter...

Страница 16: ...ion the LED Calibration should be performed before the Angle Calibration Follow the Cleaning procedure on page 21 to clean the light chamber and optic lenses 1 Tap to go to settings 2 Tab Calibration...

Страница 17: ...Universal Disk Cover Code 1719 can be used with the API AQUASPIN Keep the chamber clean and dry Gently swab LED and photodiode lenses located around the hub with a cotton swab dampened with streak fr...

Страница 18: ...able 3 To express the test result as ppm Ionized Ammonia NH4 subtract the Unionized Ammonia determined in step 2 from the Total Ammonia 10 C 15 C 20 C 25 C pH Fresh water1 Salt water2 Fresh water Salt...

Страница 19: ...nm 635 nm Display Color Capacitive Touchscreen 3 5 in 320 x 240 pixel resolution Wavelength Accuracy 2 nm Wavelength Bandwidth 10 typical Photometric Range 2 to 2 AU Photometric Precision 0 01 AU at...

Страница 20: ...anada IC Certification FCC ID WAP2001 License IC 7922A 2001 CE Europe Complies with Directive 1999 5 EC MIC Japan 005 101007 KC Korea MSIP CRM Cyp 2001 EMC EU ETSI EN 301489 1 US FCC PART 15 B CAN ICE...

Страница 21: ...horough cleaning when necessary Do not use streak free window cleaner or similar cleaners on the touchscreen REPAIRS Should it be necessary to return the meter for repair or servicing pack the meter c...

Страница 22: ...ument Failure to do so could result in personal injury or damage to the meter The API AQUASPIN should not be stored or used in a damp or excessively corrosive environment Care should be taken to preve...

Страница 23: ...ance LaMotte Company warrants this instrument to be free of defects in parts and workmanship for 2 years from the date of shipment If it should become necessary to return the instrument for service du...

Страница 24: ...pifishcare com Mars Fishcare North America Inc 50 E Hamilton St Chalfont PA 18914 Disk US Patent No 8 734 734 AP 3580 MN 07 22 20 Manufactured by LaMotte Company PO Box 329 Chestertown MD 21620 800 34...

Отзывы: