40424 - 2nd Float Offset

Used to calibrate bottom float reading (-10,364 - 10,364 mm). Differences in fluid specific gravity can change 
the level at which a float rests in the liquid. Use this parameter to match probe reading to confirmed liquid 

level.

40425 - Gain Offset (Factory Calibrated)

Used to move the centerline of the float response signal to optimize signal strength (0 - 255).

40427 - 20mA Set

Used to set the distance (in mm) which will correspond to an output of 20 mA. For Application 0 (Distance), 
this is measured from the Zero Reference. For all other applications (Level & Volumetric) this is measured 
from the bottom of the probe. See Figure 3.3.

40426 - 4mA Set

Used to set the distance (in mm) which will correspond to an output of 4 mA. For Application 0 (Distance), 
this is measured from the Zero Reference. For all other applications (Level & Volumetric) this is measured 
from the bottom of the probe. See Figure 3.3.

40428 - 4mA Cal (Factory Calibrated)

Used to calibrate the 4 mA output of the MPX-E/R2 or -E/R3.

40429 - 20mA Cal (Factory Calibrated)

Used to calibrate the 20 mA output of the MPX-E/R2 or -E/R3.

Zero

Reference

Bottom

of Probe

Application 0

Distance

0 mm

4 mA

20 mA

Application 1-11

Level or Volume

0 mm

4 mA

20 mA

Figure 3.3

Содержание MPX-E

Страница 1: ...APG R Doc 9003761 Part 200204 Rev G 06 2022 MPX Magnetostrictive Level Sensors User Manual For The MPX E MPX E Chemical and MPX R...

Страница 2: ...Installation 11 Removal Instructions 12 Chapter 3 Programming 12 Modbus Programming 12 Modbus Programming with RST 6001 and APG Modbus Software 12 4 20 mA Programming with RST 4100 and APG Modbus Sof...

Страница 3: ...rical ratings Rated 12 24 VDC 4 20 mA or 80 mA Class I Division 1 2 Groups C and D T4 Ta 85 C Ex d IIB T4 Ta 85 C Ex nA IIB T4 Ta 85 C Class I Zone 1 AEx d IIB T4 Ta 85 C Class I Zone 2 AEx na IIB T4...

Страница 4: ...material and workmanship under normal use and service of the product for 24 months For a full explanation of our Warranty please visit https www apgsensors com about us terms conditions Contact Techni...

Страница 5: ...1 63 C D 2 20 1 47 F 1 87 1 78 E Float Ref S1 6 S2 1 1 Float Ref S1 6 S2 2 27 Float Ref S1 6 S2 1 5 S2 Dead Band from Float Ref to bottom of stem Min 12 Max 153 S1 Dead Band from Zero Reference to Fl...

Страница 6: ...ad Band from Zero Reference to Float Ref S2 Dead Band from Float Ref to bottom of stem 3 4 NPT 1 2 NPT 2 13 4 94 4 15 ZERO REFERENCE Min 12 Max 153 1 91 1 88 G Float Ref S1 6 S2 2 38 Note S2 increases...

Страница 7: ...loat Ref S1 10 63 S2 6 Float Ref S1 12 25 S2 4 25 Float Ref S1 12 25 S2 4 25 Float Options Y Z L M W X S T U V I J I J U V 4 21 6 84 6 97 S1 Dead Band from Zero Reference to Float Ref S2 Dead Band fro...

Страница 8: ...0 mA 4 22 mA single 8 44 mA dual typical Protection Reverse Polarity Surge Output 4 Materials of Construction Housing Cast aluminum epoxy coated Stem MPX E 0 5 316L SS MPX E Chemical 0 5 316L SS with...

Страница 9: ...6L SS Cylindrical 0 65 SG D 316L SS Cylindrical 0 92 SG E Buna N 0 5 SG F 316 SS 3A Cylindrical 0 5 SG G Kynar Cylindrical 0 66 SG E Float 2 optional N None B 316L SS Round 0 92 SG F Mounting Type P N...

Страница 10: ...ype All Housing Die cast Aluminum IP65 Blue __ Large Housing A Small Housing D Float 1 G Kynar Cylindrical 0 66 SG E Float 2 N None F Mounting Type P NPT Plug 150 G Mounting Size 2 Size 2 Note This op...

Страница 11: ...Round 316L SS 0 60 0 94 SG M L 5 5h x 2d in Red Polyurethane 0 57 0 94 SG J I 5h x 3d in Oval Titanium 0 60 0 94 SG N None E Float 2 optional N None Y 5 5h x 3d in Blue Polyurethane 0 94 SG W 5 in Rou...

Страница 12: ...nd lug EARTH GND EARTH GND Power Supply 12 24 Vdc GND Use Shielded Cable Equivalent 120 terminating resistor internal to RST 6001 120 terminating resistor at last sensor USB to computer with APG Modbu...

Страница 13: ...on is for programming ONLY After programming sensor must be reintegrated to 4 20 mA loop for proper system operation 12 24 Vdc Out 1 Out 2 Input 1 Input 2 Com Receiver Sensor Sensor 4 20 mA Dual Float...

Страница 14: ...etc Ample space for maintenance and inspection Additional care must be taken to ensure The probe is located away from strong magnetic fields such as those produced by motors transform ers solenoid val...

Страница 15: ...th ground or ensure that tank mounting of MPX is grounded See Electrical Connections and System Wiring Diagrams pages 8 9 for Modbus and 4 20 mA wiring examples Electrical Installation If your sensor...

Страница 16: ...ly oper ate as client devices Sensor default transmission settings are 9600 Baud 8 Bits 1 Stop Bit No Parity and require a minimum delay of 300 ms between transactions See MPX E R4 Modbus Register Lis...

Страница 17: ...nsor the RST 4100 must be removed and the wiring returned to normal See 4 20 mA Loop Wiring and 4 20 mA Programming Wiring on page 9 4 20 mA Programming with RST 4100 and APG Modbus Software Modbus Re...

Страница 18: ...ffset 10 364 to 10 364 mm 40418 Pre filter 0 to 10 364 mm 40419 Noise limit 0 to 255 40420 Temperature Select 0 to 8 40421 RTD Offset 0 C 100 to 100 40422 Float Window 0 to 1 000 mm 0 1 float 40423 1s...

Страница 19: ...g 8 N A 9 Vertical Oval Tank 10 Horizontal Oval Tank 11 Strapping Chart See MPX E R Application Type Parameters pages 26 30 For Output 4 Loss of Signal Error Codes are dependent on Application Type Lo...

Страница 20: ...nsor will stop looking for float signals usually the bottom of the stem A float beyond the Maximum Distance value will not be detected 40406 Full Distance Sets the positive distance beginning from the...

Страница 21: ...ng Register 40413 necessary to automatically adjust the current reading and move the Filter Window 40415 Sample Rate Sets the update rate of the sensor between 10 1000 ms Shorter time delays allow for...

Страница 22: ...ory diagnostics only 40420 Temperature Select Selects the temperature sensor reading to be displayed in Input Register 30302 MPX E R4 sensors are limited to a single RTD sensor in the stem Only option...

Страница 23: ...d level 40425 Gain Offset Factory Calibrated Used to move the centerline of the float response signal to optimize signal strength 0 255 APG Modbus Register Lists for MPX E R2 and MPX E R3 Input Regist...

Страница 24: ...40417 Offset 10 364 to 10 364 mm 40418 Pre filter 0 to 10 364 mm 40419 Noise limit 0 to 255 40420 NA NA 40421 RTD Offset 0 C 100 to 100 40422 Float Window 0 to 1 000 mm 0 1 float 40423 1st Float Offs...

Страница 25: ...Ends 6 Spherical Tank 7 Pounds Linear Scaling 8 N A 9 Vertical Oval Tank 10 Horizontal Oval Tank 11 Strapping Chart See MPX E R Application Type Parameters pages 26 30 For the MPX E R2 and MPX E R3 th...

Страница 26: ...etostrictive wire 40410 Blanking Factory Calibrated Sets the blanking distance which is the zone from the Zero Reference of the sensor to the point from which the first signal will be valid Signals fr...

Страница 27: ...te the sensor 10 1000 ms Shorter time delays allow for quicker sensor response times to changing levels Typical setting is 200 ms Settings under 200 ms are not recommended 40412 Averaging Sets the num...

Страница 28: ...efines the physical range 0 10 364 mm of the start up pre filter window Four sample readings must be found within the Pre filter window for the MPX sensor to successfully start up This register is use...

Страница 29: ...Application 0 Distance this is measured from the Zero Reference For all other applications Level Volumetric this is measured from the bottom of the probe See Figure 3 3 40426 4mA Set Used to set the...

Страница 30: ...Units 40404 Decimal Calculated 0 3 Application 2 Volume of Standing Cylindrical Tank Hemispherical Bottom Register Function Value Range 40400 Device Address 1 to 247 40401 Units 40402 Application Type...

Страница 31: ...m 40440 40441 Length height of Cone 0 1 000 000 mm Application 4 Volume of Standing Rectangular Tank Chute Bottom Register Function Value Range 40400 Device Address 1 to 247 40401 Units 40402 Applicat...

Страница 32: ...7 Empty Distance 0 10 364 mm 0 11 278 mm 40436 40437 Tank Length 0 1 000 000 mm 40438 40439 Tank Diameter 0 1 000 000 mm 40440 40441 Radius of End Hemispheres 0 1 000 000 mm Application 6 Volume of Sp...

Страница 33: ...10 364 mm 0 11 278 mm 40436 40437 Tank Length 0 1 000 000 mm 40438 40439 Tank Depth 0 1 000 000 mm 40440 40441 Tank Width 0 1 000 000 mm Application 7 Pounds Linear Scaling Register Function Value Ran...

Страница 34: ...1 000 000 mm 40438 40439 Tank Depth 0 1 000 000 mm 40440 40441 Tank Width 0 1 000 000 mm Application 11 Strapping Chart Polynomial Values Register Function Value Range 40400 Device Address 1 to 247 40...

Страница 35: ...ment immediately Chapter 4 Maintenance General Care Should your MPX level sensor require service please contact the factory via phone email or online chat We will issue you a Return Material Authoriza...

Страница 36: ...32 Tel 1 888 525 7300 Fax 1 435 753 7490 www apgsensors com sales apgsensors com Chapter 5 Hazardous Location Installation and Certification Hazardous Location and Non Incendive Wiring Diagram...

Страница 37: ...with adjacent indicators C and US for Canada and US or with adjacent indicator US for US only or without either indicator for Canada only Issued by Adrian Zilahi Adrian Zilahi P Eng PRODUCTS CLASS 225...

Страница 38: ...s as needed f F R P S N or O g 1 1 5 2 2 5 3 3 5 4 5 6 h W S or T New mount connection i A B C D E F or G additional wetted materials include sleeves or baked on coatings and j 12 153 inches for the 1...

Страница 39: ...nth Edition Reprint with Revisions Through and Including April 15 2010 UL 1203 Explosion Proof and Dust Ignition Proof Electrical Equipment for Use in Hazardous Classified Locations Fourth Edition Rep...

Страница 40: ...Automation Products Group Inc Tel 1 888 525 7300 Fax 1 435 753 7490 www apgsensors com sales apgsensors com APG R...

Отзывы: