2020.07.24
•
REV. 2
• MASS FLOW METER OPERATING MANUAL
19
Totalizer (Optional)
TOTAL/TIMER
toggles between
totalized flow and
elapsed time as
the parameter
highlighted in
the center
M PEAK
or
V PEAK
displays the
maximum flow rate
since the last reset. Press to
select engineering units.
Displays live flow rate. Press
to select engineering units.
RESET
clears all totalized data
and immediately resets the
timer to 0. Totalization of flow
data continues immediately.
MENU/MAIN
enters the menu
system (
). From
there, press
MAIN
to exit to
the Main Display of live data.
•
M AVG
: Shows optional
totalizer averaging, which
displays average flow rate
since last reset, updated live
•
Main display shows either time
since reset, or totalized flow
Totalizer Rollover Functions
Your flow totalizer has been configured to report a maximum of 7 digits. By
default, the placement of the decimal is the same as the live flow rate. The
totalizer can be configured at the time of order for the following behaviors.
• Rollover
(default)
:
Totalizer resumes
counting from 0
as soon as the
maximum count has
been reached.
• Freeze:
Totalizer
stops counting at
max count, until it
is reset manually.
• Error:
Displays
OVR
status message when
maximum count
has been reached;
compatible with
Rollover and Freeze.
The time counter has a maximum value of 9999:59:59 (h:m:s) (416 days, 16
hours). If flow is still being totalized at that point, the timer freezes, regardless
of totalized flow settings.
Option: Collecting Totalized Flow Data
Your flow meter may have been purchased with an optional flow totalizer.
This displays the total amount of mass or volume that has flowed through
the instrument since its last reset, like a gasoline pump. You can access
the totalizer screen by pressing
TOTAL/MENU
on the Main Display.
TOTAL/ M PEAK
SLPM
TIMER +18.030
+15.45
M AVG +0:24:56 h:m:s
+24:56
+45.290 SL
MENU/
RESET
MAIN
TOTAL/ M PEAK
SLPM
TIMER +18.030
+15.45
M AVG +45.290 SL
+45.29
+0:45:56 h:m:s
MENU/
RESET
MAIN