• MASS FLOW METER OPERATING MANUAL
23
Gas Select
In most cases, your flow meter was physically
calibrated on air at the factory. Gas Select
allows you to reconfigure the flow meter to
flow a different gas without sending it back
for a physical recalibration.
To use Gas Select, simply choose a gas or
gas mix from one of the listed categories.
As soon as you press
SELECT
from the gas
listing, your flow meter will reconfigure its
calculations to flow your chosen gas. There
is no need to restart the flow meter.
Your current gas selection appears just
below the unit’s indicator on the right side
of the Main Display. In the example to the
right, the gas is set to nitrogen gas (
N2
).
PSIA
¤C
TARE
+13.59 +25.33 FLOW
+15.404
Mass Flow
SLPM
N2
+16.670 +15.404 TOTAL/
LPM
SLPM
MENU
A meter set to measure N₂
Menu
→
Basic Config
→
Gas
Category and Gas Listings Controls
•
UP/DOWN
moves the selection arrow up or down the listing of gas
categories (in the Category Menu) or gases (Gas Listing Menu).
•
PAGE
advances the view to the next page of categories.
•
BACK
returns to the Basic Configuration Menu (
).
•
SELECT
(in the Category Listing Menu) opens
a list of gases in that category.
•
SET
(in the Gas Listing Menu) immediately loads the gas
measurement properties and exits to the Main Display (
).
UP
DOWN
PAGE
Recent
>Standard
COMPOSER User Mixes
Bioreactor
Breathing
Chromatography
Fuel
Laser
BACK
SELECT
Category Listing Menu
UP
DOWN
PAGE
Air
>Ar Argon
CH4 Methane
CO Carbon Monoxide
CO2 Carbon Dioxide
C2H6 Ethane
H2 Hydrogen
He Helium
BACK
SET
Gas Listing Menu