1.  Introduction 

Thank you for choosing Apera Instruments PH60F-Z Smart Flat pH Tester. Please carefully read this 
manual before using the product.

 

 

 

Search “zentest smart” in Apple App Store or Google Play App Store to download the latest 

App for your tester. Turn on 

the Bluetooth of your phone, go to ZenTest™ App, tap 

on the 

upper right corner, then select your tester to connect. 

  For video tutorials on how to get the most out of 

ZenTest™

, please go to 

www.aperainst.de

 

  This product is designed with a two-

way control on both the tester and ZenTest™ App. Please 

refer to the functions available on each platform in the following table. This manual shows you how 

to operate the tester without connecting to a smartphone.   

 

Table 1: Funtions on 60-Z Tester and ZenTest

® 

Mobile App

 

Functions

 

60-Z Tester

 

ZenTest Mobile App

 

Display

 

LCD display

 

1. 

Basic Mode: digital display 
+calibration info

 

Swipe to 
switch 
among 
various 
modes

 

2. 

Dial Mode: digital dial 
display

 

3. 

Graph Mode: digital graph 
display

 

4. 

Table Mode: digital real time 
measurement and history display

 

Calibration

 

Press buttons to 
operate

 

Operate on smartphone following graphic guides

 

Self-
Diagnosis

 

Er1 – Er6 icons

 

Detailed problem analysis and solutions

 

Parameter 
Setup

 

Press buttons to set up 
(except for P7 and P11)

 

All parameters can be set up in Settings.

 

Alarm

 

The screen turns red 
when alarm triggered; 
cannot be setup

 

Alarm display and alarm values can be preset for each 
parameter

 

Datalogger

 

N/A

 

Manual or Auto. Datalogger; notes can be added to 
saved data

 

Data 
Output

 

N/A

 

Share data via Email

 

Содержание AI3715

Страница 1: ...1 ZenTest PH60F Z Smart Flat pH Tester pH Temp ORP Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...Storage 8 11 Parameter Setting 9 12 Technical Specifications 10 13 Icons and Functions 10 14 Probe Replacement 11 15 Troubleshooting Guide 11 16 Warranty 12 ATTENTION Water droplets are added during...

Страница 3: ...ut connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LCD display 1 Basic Mode digital display calibration info Swipe to...

Страница 4: ...on Please install batteries according to the following steps Please note direction of batteries All POSITIVE SIDES FACING UP Wrong installation of batteries will cause damage to the tester and potenti...

Страница 5: ...urn on the meter rinse the probe in distilled water Shake off excess water 1 When turned off short press to turn on the tester long press to enter parameter setting 2 In calibration mode or parameter...

Страница 6: ...after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH 4 00...

Страница 7: ...ed Please use a soft brush with soap water to clean it thoroughly Then soak it in 3M KCL overnight before performing calibration again 3 The electrode is aged used for over a year and has a much slowe...

Страница 8: ...d water and dry it Dip the probe in sample solution shake for a few seconds and allow it to stand still Get the ORP readings after appears and stays on screen 9 Probe Cleaning 1 The tester is only as...

Страница 9: ...isplayed when reading is locked 3 Auto Backlight P03 Users can set the automatic backlight time for 1 to 8 minutes For example if 3 minutes is set the backlight will turn off automatically after 3 min...

Страница 10: ...Icons and Functions pH Range 2 00 to 16 00 pH Resolution 0 01 pH Accuracy 0 01 pH 1 digit Calibration Points 1 to 3 points Auto Temperature Compensation 0 50 C 32 122 F ORP mV Range 1000 mV to 1000 mV...

Страница 11: ...ng Incorrect standard solutions Reboot tester calibrate pH 7 first then pH 4 For details refer to Section 5 2 a Poor quality standard solutions Replace with fresh and clean standard calibration soluti...

Страница 12: ...ean the probe with cleaning solution then soak it in 3M KCL soaking solution overnight Comparison with other testers test strips or drop tests To compare with other testers make sure to perform a 2 po...

Отзывы: