- 10 -

 

appear in the lower right corner of the LCD screen in 3 days to remind you to perform calibration, 
also in the ZenTest App there will be a pop-up reminder. After calibration is finished or the reminder 
setting is cancelled in the ZenTest App, the Er6 icon will disappear.

 

6) 

pH Back to Factory Default (P08) 

– Select “Yes” to recover instrument calibration to theoretical 

value.  This  function  can  be  used  when  instrument  does  not  work  well  in  calibration  or 
measurement. Calibrate and measure again after setting the instrument back to factory default. 

 

12. Technical Specifications 

13. Icons and Functions 

 

 

 
 

pH 

Range 

-2.00 to 16.00 pH 

Resolution 

0.01 pH 

Accuracy 

±0.01 pH ±1 digit 

Calibration Points 

1 to 3 points 

Auto. Temperature Compensation 

 

50˚C (32 –

 

122˚F)

 

ORP (mV) 

Range 

-1000 mV to 1000 mV 

Accuracy 

±0.2% F.S 

Temperature 

Range 

0 to 50˚C (32

-

122˚F)

 

Accuracy 

±0.5˚C

 

Calibrated points

 

 

Self-Diagnosis 

Symbol

 

Er1, Er2, Er3, Er4, Er5, Er6

 

Stable reading indicator

 

 

Waterproof Rating 

IP67, floats on water 

Reading Lock

 

HOLD

 

Power 

DC3V, AAA batteries*4 

Bluetooth Signal

 

 

Battery Life 

200 Hours 

Low power reminder

 

         

Backlight 

White: Measurement.   

Green: Calibration; Red: Alarm 

Auto. Power Off 

Automatically power off if no operation for 10 minutes

 

Dimension/Weight 

Instrument

40×40×178mm/133g; case

255×210×50mm/680g; 

Graph-3

 

LCD Display

 

Graph-4

 

pH calibration reminder

 

 

Graph-5

 

pH alarm triggered

 

 

Содержание AI3715

Страница 1: ...1 ZenTest PH60F Z Smart Flat pH Tester pH Temp ORP Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...Storage 8 11 Parameter Setting 9 12 Technical Specifications 10 13 Icons and Functions 10 14 Probe Replacement 11 15 Troubleshooting Guide 11 16 Warranty 12 ATTENTION Water droplets are added during...

Страница 3: ...ut connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LCD display 1 Basic Mode digital display calibration info Swipe to...

Страница 4: ...on Please install batteries according to the following steps Please note direction of batteries All POSITIVE SIDES FACING UP Wrong installation of batteries will cause damage to the tester and potenti...

Страница 5: ...urn on the meter rinse the probe in distilled water Shake off excess water 1 When turned off short press to turn on the tester long press to enter parameter setting 2 In calibration mode or parameter...

Страница 6: ...after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH 4 00...

Страница 7: ...ed Please use a soft brush with soap water to clean it thoroughly Then soak it in 3M KCL overnight before performing calibration again 3 The electrode is aged used for over a year and has a much slowe...

Страница 8: ...d water and dry it Dip the probe in sample solution shake for a few seconds and allow it to stand still Get the ORP readings after appears and stays on screen 9 Probe Cleaning 1 The tester is only as...

Страница 9: ...isplayed when reading is locked 3 Auto Backlight P03 Users can set the automatic backlight time for 1 to 8 minutes For example if 3 minutes is set the backlight will turn off automatically after 3 min...

Страница 10: ...Icons and Functions pH Range 2 00 to 16 00 pH Resolution 0 01 pH Accuracy 0 01 pH 1 digit Calibration Points 1 to 3 points Auto Temperature Compensation 0 50 C 32 122 F ORP mV Range 1000 mV to 1000 mV...

Страница 11: ...ng Incorrect standard solutions Reboot tester calibrate pH 7 first then pH 4 For details refer to Section 5 2 a Poor quality standard solutions Replace with fresh and clean standard calibration soluti...

Страница 12: ...ean the probe with cleaning solution then soak it in 3M KCL soaking solution overnight Comparison with other testers test strips or drop tests To compare with other testers make sure to perform a 2 po...

Отзывы: