background image

 

Making a Soil Slurry 

6.1 

Remove about 2 inches (5 

cm) of the top layer soil and collect different locations’ soil 

samples at approximately 6 inches (15 cm) deep. 

6.2 

Thoroughly mix all collected soil. 

6.3 

Ideally, dry the soil in the air or bake in an oven at 104 

˚F / 40 ˚C. 

6.4 

Weigh out 20 g of the mixed soil sample into a glass jar and add 100 g of  distilled or 

deionized water in it. 

6.5 

Shake well for 5 minutes or use a magnetic stirrer to automatically stir for 15 minutes. 

Leave overnight.  

6.6 

Shake or stir again next morning, then allow it to settle for 15-30 minutes. 

6.7 

Power on the tester; Remove the probe cap; Rinse the probe in the water  bucket and 

shake dry. 

6.8 

Submerge  the  probe  into  the  soil  solution,  shake  for  a  few  seconds,  and  wait  for  the 

reading to fully stabilize. Then record the reading. 

6.9 

After  measurement,  put  the  probe  back  in  the  storage  cap  and  soak  in  the  3M  KCl 

solution. 

 

Other Functions

 

7.1 

If necessary, you can manually hold (lock) the reading by short pressing 

. Press it 

again to cancel the hold. 

7.2 

Long press 

 to switch temp. unit between °F and °C. 

7.3 

 

 

 

 

 

 

Out of range reminder for pH  

Out of range reminder for temperature 

 

7.4 

The tester will automatically power off if there is no operation within 

10  minutes.  If  you  want  to  turn  off  the  Auto.  Power  Off  function, 

power off the tester, and then hold 

 for 5 seconds until you see 

Auto  off.  Then  it  will  power  on  and  go  to  measurement  mode 

automatically.  

 

Содержание 28APE190350/AI31020

Страница 1: ...GroStar GS2 Premium Soil pH Pen Tester User Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...for the horticultural market Since 1991 Apera instruments has been dedicated to providing hi tech accurate lab grade instruments and sensors GroStar s intelligent design reduces the guesswork so you...

Страница 3: ...the backlight Start calibration In calibration mode press to cancel calibration Switch between F and C Swiss LabSen Spear Soil pH Probe Backed by Apera s proprietary sensor technology from Switzerland...

Страница 4: ...H Calibration 3 1 Power on the pen and remove the probe cap Always calibrate 7 00 pH first 3 2 Rinse the probe with clean water and shake off excess water then submerge it in the 7 00 pH standard buff...

Страница 5: ...f a discrepancy is found then it s time to calibrate again 3 6 If the calibration fails the screen will turn red For details see Section 13 Troubleshooting Guide Always perform at least a 2 point pH c...

Страница 6: ...the white PTFE junction ring using the cleaning brush Avoid brushing the spear glass tip After cleaning shake off excess water For details of probe cleaning refer to Section 8 5 6 Repeat Step 5 2 to...

Страница 7: ...cap Rinse the probe in the water bucket and shake dry 6 8 Submerge the probe into the soil solution shake for a few seconds and wait for the reading to fully stabilize Then record the reading 6 9 Aft...

Страница 8: ...robe in Apera s cleaning solution or detergent water for 30 minutes Then use a soft brush to remove the contaminants Afterwards soak the probe in 3M KCL soaking solution for 1 hour Rinse it off then r...

Страница 9: ...w probe make sure to align the connector s position properly and screw on the probe ring pH probes don t last forever Every pH probe will eventually age and fail even if you don t use it that often Th...

Страница 10: ...fic instrument manufacturers Contaminated probe Clean the probe with Apera s cleaning solution or detergent water Aged probe Replace the probe Dried out probe Soak the probe in the 3M KCl soaking solu...

Страница 11: ...or consistent Contaminated probe After each measurement make sure to use a soft brush with water to clean off all the dirt on the probe especially the white PTFE junction ring Avoid wiping or rubbing...

Страница 12: ...10 pH 10 01 solution sold separately Unit pH F C Power supply 4 AAA alkaline batteries Backlight White measurement Green calibration Red errors Reading hold Manual Warranty Two years for the instrumen...

Страница 13: ...receipt for Product showing model number payment and date of purchase This warranty is non transferable and terminates if the original purchaser consumer sells or transfers the Product a third party...

Страница 14: ...horized distributor or reseller How Do You Get Service To begin a warranty claim you must return the Product to the point of purchase with valid proof of purchase Limitation of Liability Acknowledgmen...

Страница 15: ...MAINING PROVISIONS OF THE WARRANTY SHALL REMAIN IN FULL FORCE AND EFFECT Governing Law Authority This warranty is governed by the laws of the state of country where Product is purchased without regard...

Отзывы: