3-Color LCD Screen 

 

White for measurement mode 

 

Green for calibration mode 

 

Red for calibration error 

 stands for stabilized reading 

L/M/H 

stands 

for 

successful 

calibration: 

is pH 4, 

is pH 7 

pH 

is 

pH  10.  The  icon(s)  will  disappear 
after 30 days to remind you for re-
calibration 

Probe Cap Ring 

Loosen  the  probe  cap  ring  before 

pulling out the probe cap  

 

Short Press 

Long 

Press 

 

1. Power on 

2. Manual hold 

Press again to 

cancel hold 

Power off 

 

In measurement 

mode, press to 

turn on/off the 

backlight. 

Start 

calibration 

 

In calibration 

mode, press to 

cancel calibration 

Switch 

between 

°F and °C 

Swiss LabSen® Spear Soil pH Probe 

Backed by Apera's proprietary sensor 

technology  from  Switzerland,  this  spear 

pH  probe  is  specifically  designed  for 

making  soil  pH  test  easier  and  more 

reliable. 

 

Impact-resistant spear glass tip, 10 

times stronger than conventional glass 

bulb 

 

PTFE ring prevents a clogged junction 

caused by soil particle 

 

Long-life reference system ensures 

high durability

 

Probe Cap 

 

The probe must be soaked in 3M 
KCl soaking solution when not in 
use 

 

Fill in 3M KCl soaking solution to 
about 1/3 of the probe cap, store 
the probe in it, then twist on the 
cap ring

 

Содержание 28APE190350/AI31020

Страница 1: ...GroStar GS2 Premium Soil pH Pen Tester User Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...for the horticultural market Since 1991 Apera instruments has been dedicated to providing hi tech accurate lab grade instruments and sensors GroStar s intelligent design reduces the guesswork so you...

Страница 3: ...the backlight Start calibration In calibration mode press to cancel calibration Switch between F and C Swiss LabSen Spear Soil pH Probe Backed by Apera s proprietary sensor technology from Switzerland...

Страница 4: ...H Calibration 3 1 Power on the pen and remove the probe cap Always calibrate 7 00 pH first 3 2 Rinse the probe with clean water and shake off excess water then submerge it in the 7 00 pH standard buff...

Страница 5: ...f a discrepancy is found then it s time to calibrate again 3 6 If the calibration fails the screen will turn red For details see Section 13 Troubleshooting Guide Always perform at least a 2 point pH c...

Страница 6: ...the white PTFE junction ring using the cleaning brush Avoid brushing the spear glass tip After cleaning shake off excess water For details of probe cleaning refer to Section 8 5 6 Repeat Step 5 2 to...

Страница 7: ...cap Rinse the probe in the water bucket and shake dry 6 8 Submerge the probe into the soil solution shake for a few seconds and wait for the reading to fully stabilize Then record the reading 6 9 Aft...

Страница 8: ...robe in Apera s cleaning solution or detergent water for 30 minutes Then use a soft brush to remove the contaminants Afterwards soak the probe in 3M KCL soaking solution for 1 hour Rinse it off then r...

Страница 9: ...w probe make sure to align the connector s position properly and screw on the probe ring pH probes don t last forever Every pH probe will eventually age and fail even if you don t use it that often Th...

Страница 10: ...fic instrument manufacturers Contaminated probe Clean the probe with Apera s cleaning solution or detergent water Aged probe Replace the probe Dried out probe Soak the probe in the 3M KCl soaking solu...

Страница 11: ...or consistent Contaminated probe After each measurement make sure to use a soft brush with water to clean off all the dirt on the probe especially the white PTFE junction ring Avoid wiping or rubbing...

Страница 12: ...10 pH 10 01 solution sold separately Unit pH F C Power supply 4 AAA alkaline batteries Backlight White measurement Green calibration Red errors Reading hold Manual Warranty Two years for the instrumen...

Страница 13: ...receipt for Product showing model number payment and date of purchase This warranty is non transferable and terminates if the original purchaser consumer sells or transfers the Product a third party...

Страница 14: ...horized distributor or reseller How Do You Get Service To begin a warranty claim you must return the Product to the point of purchase with valid proof of purchase Limitation of Liability Acknowledgmen...

Страница 15: ...MAINING PROVISIONS OF THE WARRANTY SHALL REMAIN IN FULL FORCE AND EFFECT Governing Law Authority This warranty is governed by the laws of the state of country where Product is purchased without regard...

Отзывы: