Apel BR-5000N Скачать руководство пользователя страница 5

BR5N-1-OMA-103a 

4

1.3 

Important  Safety  Instruction

 

 
When utilizing your 

BR-5000N

, always take basic safety precautions and warnings, in 

order to avoid damage to your unit, property and/or personal injury, which may result 
from misusage, exposure to fire or electric shock.    Important warnings and safety 
precautions are as follows:   
 
1.  To ensure accuracy of the measurement, always warm-up the instrument for at 

least 15 minutes, prior to operation. 

 

 

 

 

 

 

2.  Never cover the unit when it is turned-on.    Doing so will prevent the electrical 

components from being cooled properly and cause damage. 

 

 
3.  Unplug the 

BR-5000N

 from the wall outlet prior to cleaning.    Do not use organic 

liquid solutions such as alcohol or aerosol cleaners.    Use a soft dry cloth for 
cleaning. 

 

 
4.  Do not directly expose your unit to water or liquid.    Avoid spills, which could 

damage electronic components.   

 

 

 

 

 

 

 

5.  Always use appropriate capillary tube in specification, when obtaining bilirubin’s 

data or using standard solution.  

 

 

 

 

 
6.  Place this product on stable cart, stand, or table.    

 

 

 

 

7.  If stored in non-room temperature setting, allow the unit to reach room 

temperature before turning on.  Do not place this instrument in an environment 
where there is severe temperature variation. 

 

 

 

 

 

 

 

 

8.  Place the unit away from dust, excessive moisture, or corrosive chemicals.   

Cover the instrument when not in use to protect electrical components from dust.
 

 

 

9.  Do not dismantle or modify the instrument.    Damage or performance problems 

resulting from disassembly or alteration of the unit, will not be covered under the 
warranty. Contact your supplier when maintenance, service or repair work are 
required.   

   
10. When instrument is not in use, make sure the power switch has been turned-off 

and unplug the power cord from the electrical outlet.    Do not pull the cord, but 
hold the receptacle of the plug and pull it. 

 
11. Only use light bulb, capillary tubes and power cord that come with this instrument.   

Such replacement parts are available through your supplier. 

 
12. Do not overload wall outlets with extension cords as this can result in the risk of 

fire or electric shock. Attaching multiple loads to the same wall outlet is 
dangerous and may cause improper operation of BR-5000N. 

 

Содержание BR-5000N

Страница 1: ...please read this manual carefully and retain it for future reference Failure to read understand and follow the instructions in this manual may result in damage to the unit injury to operating personne...

Страница 2: ...or 7 Section 3 Operation procedure 3 1 Preparation 8 3 2 Taking blood sample 8 3 3 Measurement of the specimen 10 3 4 Setting the name or ID number of the test 12 3 5 Other functions 12 3 6 How to use...

Страница 3: ...oped with the purpose of enabling the measurement of total bilirubin in serum of neonate while compensating for hemolysis turbidity and other components of obstruction With the BR 5000N an alarm is di...

Страница 4: ...to computer by serial interface The measurement data can be sent to computer through RS 232C interface 7 mg and mol indication One of mg and mol indication can be selected by the control panel 8 Easy...

Страница 5: ...on stable cart stand or table 7 If stored in non room temperature setting allow the unit to reach room temperature before turning on Do not place this instrument in an environment where there is seve...

Страница 6: ...refer to your supplier under the following conditions When the power cord or plug is damaged and or frayed If the unit has been directly exposed to water or liquid If the instrument displays unusual...

Страница 7: ...nance repair or service work In the event of missing parts defects damages or other problems immediately contact your supplier Make sure that your carton contains the following items Main Unit Quantit...

Страница 8: ...mp 8 Fuse holder 3 Control panel 9 AC Power socket 4 Serum gauge 10 Printer Interface Connector D sub25p female 5 Capillary tube holder 11 RS 232C Interface Connector D sub9p male 6 Serial number labe...

Страница 9: ...le from neonates warm the tip of the big toe the heel of the foot or wherever blood is to be drawn to get the blood flowing well Taking the blood sample within 2 hours after suckling or taking meal ma...

Страница 10: ...serum part A serum gauge is provided on the surface of BR 5000N Fig 2 1 Fig 2 2 Fig 2 3 Fig Fig 4 1 Fig 4 2 Top of the big toe or heel of foot Capillary tube Capillary tube Blood Blood Capillary tube...

Страница 11: ...ys can be stored and can be controled with them Measured data are stored up to 100 until the power switch is turned off For example if the samples of 10 babys are measured 10 data per baby can be stor...

Страница 12: ...ng of the printer 1 Auto print at each measurement Turn on the Auto printer SET 5 Refer to 3 5 3 2 Manual operation of printing right after the measurement Press PRINT after a data is measured 3 Print...

Страница 13: ...urement SET 0 Standard CONC 10 0 30 0 mg SET 4 Printer Use 1 on 2 off SET 5 Auto Print 1 on 2 off SET 6 Comm RS 232C 1 on 2 off SET 7 Baud rate 1 9600 2 19200 3 34800 bps SET 8 RS 232C mode d Data bit...

Страница 14: ...ixed and memorized Display comes back to Ready without memorizing the data when EXIT key is pressed Even when the Auto print is off the measured data can be printed out after the measurement with PRIN...

Страница 15: ...ion is fixed and memorized Display comes back to Ready without memorizing the data when EXIT key is pressed 3 5 7 Deletion of data You can delete the stored data for the designated ID_No And SN _ come...

Страница 16: ...the thumb touching the right side of the gauge 3 Lift up the capillary tube as it is and insert it into the capillary holder until the thumb touchs the top of the capillary holder 4 Then the slit of...

Страница 17: ...ed To use with BR 5000N Empty the standard solution into a larger container Add 6m of distilled or deionized water and dissolve the standard solution again Assay value 1 3 concentration standard is ob...

Страница 18: ...ower cord into electrical wall outlet and the unit securely LCD display and or LED does not glow when the power switch is turned on 2 DC power suply has some troubles 3 Certain cables inside are not s...

Страница 19: ...Disconnect the lamp connector fig7 4 Turn the light bulb counterclockwise toward the bulb and remove it Do not touch the bulb by your fingers directly but hold it with a cloth fig8 5 Replace the bulb...

Страница 20: ...ntainer Hematocrit capillary tube Heparinized Light source 5V 0 75Atungsten lump Detector Silicon photocell Display LCD Display 16 rows 2 lines Power requirements 100 240 50 60 10VA Dimensions 280mm 2...

Отзывы: