BR5N-1-OMA-103a 

11

5.    The  sample  No.(

 SN.XXX

 SN.XXX

 SN.XXX

 SN.XXX

)  shown  on  the  LCD  display  is  automatically   

incre1)  for  the  next  measurement.     

 

6.  When the total measurement numbers reach 100, and one more measurement is 

tried to be executed, the alarm buzzer sounds and   

the LCD display shows “

* Memory Full *

* Memory Full *

* Memory Full *

* Memory Full *

”   

as shown right figure. 
 
Press “SET” or “PRINT” to print out the whole data stored in the instrument. 
If you want to skip the print, Press “EXIT”. 
After the print or when the EXIT is pressed, LCD display   
shows “Clear all data?” Press “SET” for clearning all data,   
or press “EXIT” to skip the procedure. 
 
* Any measurement can not be done unless stored cata are not cleared, when 

*Memory Full*

 is shown. 

 

3.3.3    Print-out of measured data 

       

The printer can be connected to BR-5000N and the measured data can be   
printed out. 

        Connect the instrument and the printer with the printer cable, and turn on the 

switch of the printer. 

        Turn on “PRINTER USE” of “SET 4” which is described in the section 3.5.2 

“The setting of the printer”. 

1.  Auto print at each measurement. 

Turn on the “Auto printer (SET 5)”. (Refer to 3.5.3) 

2.  Manual operation of printing right after the measurement. 

Press “PRINT” after a data is measured. 

3.  Printing of whole data with designated ID_No. 

Set the ID_No. and Press “EXIT” key then press   
“PRINT” key. 

4.  Printing of the whole data. 

When the memory area is fully stored, the whole stored data are printed. 
Refer to 3.3.2 “Memory Full” 
 

3.3.4    Data communication function 

            Display  and  collection  of  data  on  the  Computer  can  be  done  through   

the  RS-232C  cable  from  the  instrument. 
Turn  on  “COMM(RS-232C),  SET  6”  as  described  in  the  section  3.5.4,   
if  the  communication  is  used. 
Adjust  the  baud  rate  and  the  protocol  to  the  computer. 

 

       

   

Data format for output. 

       2

22

  

 1

11

1, 

, S

SS

  

 A

AA

  

 M

MM

  

 P

PP

  

 

 

  

 0

00

  

 0

00

  

 5

55

5, 

, 0

00

  

 0

00

  

 1

11

  

  ,   

,   

,   

,   2

22

  

 1

11

  

 .

..

  

 3

33

3 0D 0A 

      ID_No.  ID_Code, Name    Sample No.   Data     CR LF 

 

* Memory Full 

* Memory Full 

* Memory Full 

* Memory Full * 

Print all data ?

Print all data ?

Print all data ?

Print all data ?

* Memory Full * 

* Memory Full * 

* Memory Full * 

* Memory Full * 
Clear all data ?

Clear all data ?

Clear all data ?

Clear all data ?

No.12

No.12

No.12

No.12   

   

   

    [Ready] 

[Ready] 

[Ready] 

[Ready] 

                

                

                

                

 [Print]  No.12 

 [Print]  No.12 

 [Print]  No.12 

 [Print]  No.12 
                

                

                

                

Содержание BR-5000N

Страница 1: ...please read this manual carefully and retain it for future reference Failure to read understand and follow the instructions in this manual may result in damage to the unit injury to operating personne...

Страница 2: ...or 7 Section 3 Operation procedure 3 1 Preparation 8 3 2 Taking blood sample 8 3 3 Measurement of the specimen 10 3 4 Setting the name or ID number of the test 12 3 5 Other functions 12 3 6 How to use...

Страница 3: ...oped with the purpose of enabling the measurement of total bilirubin in serum of neonate while compensating for hemolysis turbidity and other components of obstruction With the BR 5000N an alarm is di...

Страница 4: ...to computer by serial interface The measurement data can be sent to computer through RS 232C interface 7 mg and mol indication One of mg and mol indication can be selected by the control panel 8 Easy...

Страница 5: ...on stable cart stand or table 7 If stored in non room temperature setting allow the unit to reach room temperature before turning on Do not place this instrument in an environment where there is seve...

Страница 6: ...refer to your supplier under the following conditions When the power cord or plug is damaged and or frayed If the unit has been directly exposed to water or liquid If the instrument displays unusual...

Страница 7: ...nance repair or service work In the event of missing parts defects damages or other problems immediately contact your supplier Make sure that your carton contains the following items Main Unit Quantit...

Страница 8: ...mp 8 Fuse holder 3 Control panel 9 AC Power socket 4 Serum gauge 10 Printer Interface Connector D sub25p female 5 Capillary tube holder 11 RS 232C Interface Connector D sub9p male 6 Serial number labe...

Страница 9: ...le from neonates warm the tip of the big toe the heel of the foot or wherever blood is to be drawn to get the blood flowing well Taking the blood sample within 2 hours after suckling or taking meal ma...

Страница 10: ...serum part A serum gauge is provided on the surface of BR 5000N Fig 2 1 Fig 2 2 Fig 2 3 Fig Fig 4 1 Fig 4 2 Top of the big toe or heel of foot Capillary tube Capillary tube Blood Blood Capillary tube...

Страница 11: ...ys can be stored and can be controled with them Measured data are stored up to 100 until the power switch is turned off For example if the samples of 10 babys are measured 10 data per baby can be stor...

Страница 12: ...ng of the printer 1 Auto print at each measurement Turn on the Auto printer SET 5 Refer to 3 5 3 2 Manual operation of printing right after the measurement Press PRINT after a data is measured 3 Print...

Страница 13: ...urement SET 0 Standard CONC 10 0 30 0 mg SET 4 Printer Use 1 on 2 off SET 5 Auto Print 1 on 2 off SET 6 Comm RS 232C 1 on 2 off SET 7 Baud rate 1 9600 2 19200 3 34800 bps SET 8 RS 232C mode d Data bit...

Страница 14: ...ixed and memorized Display comes back to Ready without memorizing the data when EXIT key is pressed Even when the Auto print is off the measured data can be printed out after the measurement with PRIN...

Страница 15: ...ion is fixed and memorized Display comes back to Ready without memorizing the data when EXIT key is pressed 3 5 7 Deletion of data You can delete the stored data for the designated ID_No And SN _ come...

Страница 16: ...the thumb touching the right side of the gauge 3 Lift up the capillary tube as it is and insert it into the capillary holder until the thumb touchs the top of the capillary holder 4 Then the slit of...

Страница 17: ...ed To use with BR 5000N Empty the standard solution into a larger container Add 6m of distilled or deionized water and dissolve the standard solution again Assay value 1 3 concentration standard is ob...

Страница 18: ...ower cord into electrical wall outlet and the unit securely LCD display and or LED does not glow when the power switch is turned on 2 DC power suply has some troubles 3 Certain cables inside are not s...

Страница 19: ...Disconnect the lamp connector fig7 4 Turn the light bulb counterclockwise toward the bulb and remove it Do not touch the bulb by your fingers directly but hold it with a cloth fig8 5 Replace the bulb...

Страница 20: ...ntainer Hematocrit capillary tube Heparinized Light source 5V 0 75Atungsten lump Detector Silicon photocell Display LCD Display 16 rows 2 lines Power requirements 100 240 50 60 10VA Dimensions 280mm 2...

Отзывы: