Section 4 – Maintenance and inspection

 

 

4.1

 

  

 

Adjustment with standard solution 

        The standard value that is set in the equipment will fluctuate corresponding to         
        deterioration of the light bulb,    so to keep the accuracy of the instrument,     
        confirm the value every 1 month (assuming 2 hours of usage/day) or     
        after the light bulb is replaced or the instrument is not used for a long interval. 
      *Standard solution to be used for the adjustment must be diluted precisely.       
        Otherwise, measured value could have error. 
 

4.1.1 The procedure of adjustment 

 
1.  Turn on the power switch and wait until the LCD display shows “Ready”. 
 
2.  Adjust “ZERO” according to the procedure of 3.1, 3-6 . 
 
3.    Prepare standard solution with "Hi Level Check-BIL", manufactured by 

International Reagents Corp., Japan or any substitute standard solution.       

 
 

[Prepare original standard solution] 

  Remove cap and stopper of Hi Level Check-BIL. Reconstitute each vial with 3m

   

distilled or deionized water of low bacterial count. Replace the stopper. Leave it as it 
is for 5 minutes, then mix well by swirling. A high concentration standard is 
obtained. 

 
 

[To use with BR-5000N] 

  Empty the standard solution into a larger container. Add 6m

of distilled or 

deionized water and dissolve the standard solution again.    Assay value 

×

  1/3 

concentration standard is obtained. This is the specified value to be used with 
BR-5000N. 
 
* The assay value of Hi Level Check-BIL is specified per each production Lot.     
    Confirm the value in note enclosed with Hi Level Check-BIL, e.g. assay value for     
    Lot No. 7324 is 54.0mg/

.    Specified value is 54.0 

×

1/3 = 18.0mg/

 
4.  Set the standard vbalue of the standard solution according to the section 3.5.1. 
 
5.  Fill standard solution into a capillary tube (blue line)   

and set it in the capillary holder, then press the “CAL” key   
not less than 2 second. 

 
6.  The adjustment will end if the display shows “Measure End”. 
 

If “Error BBK = x.xxx” is shown, confirm the standard   
solution and the procedure of calibration, and calibrate it again.   
 
* Measure the standard solution as the normal sample,   

then confirm if the measured value falls within 

±

0.5 of   

the calibrated value. 

[Auto CAL]      

[Auto CAL]      

[Auto CAL]      

[Auto CAL]      
 Measure Start  

 Measure Start  

 Measure Start  

 Measure Start  

[Auto CAL]      

[Auto CAL]      

[Auto CAL]      

[Auto CAL]      
 Measure End    

 Measure End    

 Measure End    

 Measure End    

[Auto CAL]  20.0

[Auto CAL]  20.0

[Auto CAL]  20.0

[Auto CAL]  20.0
 

  

 Error BBK=0.000

Error BBK=0.000

Error BBK=0.000

Error BBK=0.000

Содержание BR-5000N

Страница 1: ...please read this manual carefully and retain it for future reference Failure to read understand and follow the instructions in this manual may result in damage to the unit injury to operating personne...

Страница 2: ...or 7 Section 3 Operation procedure 3 1 Preparation 8 3 2 Taking blood sample 8 3 3 Measurement of the specimen 10 3 4 Setting the name or ID number of the test 12 3 5 Other functions 12 3 6 How to use...

Страница 3: ...oped with the purpose of enabling the measurement of total bilirubin in serum of neonate while compensating for hemolysis turbidity and other components of obstruction With the BR 5000N an alarm is di...

Страница 4: ...to computer by serial interface The measurement data can be sent to computer through RS 232C interface 7 mg and mol indication One of mg and mol indication can be selected by the control panel 8 Easy...

Страница 5: ...on stable cart stand or table 7 If stored in non room temperature setting allow the unit to reach room temperature before turning on Do not place this instrument in an environment where there is seve...

Страница 6: ...refer to your supplier under the following conditions When the power cord or plug is damaged and or frayed If the unit has been directly exposed to water or liquid If the instrument displays unusual...

Страница 7: ...nance repair or service work In the event of missing parts defects damages or other problems immediately contact your supplier Make sure that your carton contains the following items Main Unit Quantit...

Страница 8: ...mp 8 Fuse holder 3 Control panel 9 AC Power socket 4 Serum gauge 10 Printer Interface Connector D sub25p female 5 Capillary tube holder 11 RS 232C Interface Connector D sub9p male 6 Serial number labe...

Страница 9: ...le from neonates warm the tip of the big toe the heel of the foot or wherever blood is to be drawn to get the blood flowing well Taking the blood sample within 2 hours after suckling or taking meal ma...

Страница 10: ...serum part A serum gauge is provided on the surface of BR 5000N Fig 2 1 Fig 2 2 Fig 2 3 Fig Fig 4 1 Fig 4 2 Top of the big toe or heel of foot Capillary tube Capillary tube Blood Blood Capillary tube...

Страница 11: ...ys can be stored and can be controled with them Measured data are stored up to 100 until the power switch is turned off For example if the samples of 10 babys are measured 10 data per baby can be stor...

Страница 12: ...ng of the printer 1 Auto print at each measurement Turn on the Auto printer SET 5 Refer to 3 5 3 2 Manual operation of printing right after the measurement Press PRINT after a data is measured 3 Print...

Страница 13: ...urement SET 0 Standard CONC 10 0 30 0 mg SET 4 Printer Use 1 on 2 off SET 5 Auto Print 1 on 2 off SET 6 Comm RS 232C 1 on 2 off SET 7 Baud rate 1 9600 2 19200 3 34800 bps SET 8 RS 232C mode d Data bit...

Страница 14: ...ixed and memorized Display comes back to Ready without memorizing the data when EXIT key is pressed Even when the Auto print is off the measured data can be printed out after the measurement with PRIN...

Страница 15: ...ion is fixed and memorized Display comes back to Ready without memorizing the data when EXIT key is pressed 3 5 7 Deletion of data You can delete the stored data for the designated ID_No And SN _ come...

Страница 16: ...the thumb touching the right side of the gauge 3 Lift up the capillary tube as it is and insert it into the capillary holder until the thumb touchs the top of the capillary holder 4 Then the slit of...

Страница 17: ...ed To use with BR 5000N Empty the standard solution into a larger container Add 6m of distilled or deionized water and dissolve the standard solution again Assay value 1 3 concentration standard is ob...

Страница 18: ...ower cord into electrical wall outlet and the unit securely LCD display and or LED does not glow when the power switch is turned on 2 DC power suply has some troubles 3 Certain cables inside are not s...

Страница 19: ...Disconnect the lamp connector fig7 4 Turn the light bulb counterclockwise toward the bulb and remove it Do not touch the bulb by your fingers directly but hold it with a cloth fig8 5 Replace the bulb...

Страница 20: ...ntainer Hematocrit capillary tube Heparinized Light source 5V 0 75Atungsten lump Detector Silicon photocell Display LCD Display 16 rows 2 lines Power requirements 100 240 50 60 10VA Dimensions 280mm 2...

Отзывы: