TECHNICAL SPECIFICATIONS
154 / 164 NTA47227-300A8
CALYS 50 / 75 / 100
- Range 10V : < (6 ppm R + 1 ppm Cal) /°C,
- Range 50V : < (5 ppm R + 2 ppm Cal) /°C.
Use the absolute value of the value measured (|L|) to calculate the accuracy.
D.1.8
DC Current (CALYS 100)
Range
Resolution
Accuracy / 1 year
Notes
±50 mA
1 µA
0.012% R + 2 µA
Rin < 25
Ω
4-20 mA
1 µA
0.012% R + 2 µA
Rin < 25
Ω
0-20 mA
1 µA
0.012% R + 2 µA
Rin < 25
Ω
Temperature coefficient: < (10 ppm R + 2 ppm Range)/°C from -10°C to +18°C and from +28°C to +50 °C.
Loop power supply = +24 V typ. (30 V max)
HART compatibility: input impedance Rin = 280
Ω
.
Linear or square law display scale.
Use the absolute value of the value measured (|L|) to calculate the accuracy.
D.1.9
Resistance (CALYS 100)
Range
Resolution
Accuracy / 1 year
Notes
400
Ω
1 m
Ω
0.010% R + 10 m
Ω
Measurement current = 0.25 mA
4000
Ω
10 m
Ω
0.010% R + 100 m
Ω
Measurement current = 0.25 mA
Temperature coefficient: < (6 ppm R + 1 ppm Range)/°C from -10°C to +18°C and from +28°C to +50 °C.
Automatic detection of connection scheme: 2 wire, 3-wire or 4-wire.
For 2 wire connection, the measurement includes the resistance of the line.
For 3 wire connection, add the out-of-balance of the line resistances.
Open circuit terminal voltage < 10 V.
Continuity test:
Open circuit for R > 1000
Ω.
Closed circuit for R < 1000
Ω.