GETTING STARTED
108 / 164 NTA47227-300A8
CALYS 50 / 75 / 100
B.2.1
Measuring DC voltage
Display the MEASUREMENT CONFIGURATION dialogue box:
Select the Vdc measurement function then the correct measurement range using the function and navigation
keys.
Confirm with ENTER.
The following ranges are available:
Range
100 mV
1 V
10 V
50 V
Auto
Resolution
1
µ
V
10
µ
V
100
µ
V
1 mV
---
Input impedance
> 1 G
Ω
> 1 G
Ω
10 M
Ω
(1)
10 M
Ω
(1)
---
(1) WEM41012-00B
10 M
Ω
; WEM41012-00A
1 M
Ω
.
The DC voltage to be measured is connected between the V and COM terminals.