
Simulation modes
C
NTA47227-300A8
135 / 164
C.1.7
Synthesizer mode
The synthesizer function is used:
to store up to 100 transmission values in permanent memory,
to recall and transmit the contents of these memories manually or automatically.
When this mode is active, the
icon appears in the transmission window.
The
key is used to launch generation of values in increasing order and the
function key is used to launch
generation of values in decreasing order.
The default parameters of this mode are displayed in the left hand side of the transmission window.
The parameters of the synthesizer mode are:
First point no.: number of the first point in a cycle
Last point no.: number of the last point in a cycle
T: the duration for which a point is transmitted.
Nbr: the number of polling cycles
Delay: delay between launch and transmission of the first point.
To change these parameters, use the Configuration
Synthesizer…
Parameters… menus.
The number of the first point may be higher than that of the last point. All points between the first and last are generated.
To edit the points to be synthesized, use the Configuration
Synthesizer…
Points…. Menus.