User Manual fluidlab R-300
Edition 202104
53
Vers.: 1.2
Figure 36:
Checking the sensor status.
6.5.2
Refresh Sensor Data
Detecting the correct orientation of the sample carrier before the actual measurement
is a crucial factor in the correct processing of the microscopic images that are recorded.
The associated sensors must always be recalibrated if no detection takes place, even if
the sample carrier appears to be positioned correctly.
1.
Open the settings menu of the
fluidlab
R-300 by tapping the menu icon
( ).
2.
Tap on the menu item “Maintenance”.
3.
Tap on the menu item “Refresh
Sensor
” (see
Figure 35).
4.
Tap on “Start
sensor refresh
” (
Figure 37).
5.
The device collects the new sensor data.
6.
The end of the measurement is indicated by “OK”.
Warning!
The device must be positioned on a flat surface, without a sample carrier but with an adapter
and a closed slider.
Figure 37
:
Refresh Sensor.