CCU 100 Electrical Data (for use with DHS 1100)
Ambient conditions
Disposal
The DHS 1100 with CCU 100 Combined Control Unit Unit is CE-compliant according to
:
• Electromagnetic compatibility (2014/30/EU, OJ L 96/79 of 29.3.2014)
Applied standards:
EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use - EMC requirements -
Part 1: General requirements
The product is classified as a class B equipment and is not intended for the use in industrial area.
• Low Voltage Directive (2014/35/EU, OJ L 96/357 of 29.3.2014):
Applied standards:
EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control and laboratory
use - Part 1: General requirements
IEC 61010-2-010: 2014 Safety requirements for electrical equipment for measurement, control and labora-
tory use - Part 2: Particular requirements for laboratory equipment for the heating of materials
Voltage (mains):
AC 100 - 240 V
Frequency (mains):
50 ... 60 Hz
Power consumption:
max. 594 VA
Mains fuses:
2 x T 6.3 A H 5 x 20 mm (ceramic tube)
Overvoltage category:
II according to EN 61010-1
Voltage output:
max. DC 48 V
Current output:
max. DC 11.5 A
Heater fuse:
15 A (Blade Fuse)
Ambient temperature:
5 °C to 35 °C
Atmospheric humidity:
max. 80 % relative, not condensing
Pollution degree:
2 according to EN 61010
Maximum operating altitude:
3000 m above sea level
The DHS 1100 and the CCU 100 have been designed for INDOOR USE ONLY!
Protect the instruments from moisture!
The instrument should be disposed of in accordance with the local regulations.