Anthogyr Xpert Unit Скачать руководство пользователя страница 4

5. DESCRIPTION  

 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 
 

N° 

Name 

Peristaltic pump 

ON/OFF Button (I/O) 

Power socket 220V / 110V 

Touch screen 

Connector for micromotor cable 

Front face  

N° 

Name 

 

N° 

Name 

Wireless pedal 

 

Micromotor 

Cabinet 

 

Bracke 

Irrigation line 

 

Micromotor protective casing 

Charging adapter 

 

Front face of protection 

Power cord 

 

Clips for irrigation line 

Micromotor cord 

 

Micromotor support 

Rear face  

a

®

 

EN - 

USER GUIDE

 

Fr

an

çais

 

Engli

sh

  

D

eu

tsc

h

 

Esp

ol

 

Ita

li

an

o

 

Portu

gu

ês

 

Ne

d

erl

an

d

中国的

 

  

   رع

 

 

 

 

It is imperative to familiarise yourself with all the safety instructions before installation and use of the device (see paragraph "2 - 
Safety").  
Refer  to  the  usage  instructions  supplied  with  the  micromotor  and/or  micrometer  cable  to  familiarise  yourself  with  all  safety, 
maintenance and usage aspects. 

 

6.1. MAINS CONNECTION  

Use only the power cable supplied by the manufacturer.  

Do not forget to verify that the supply voltage is exactly same as the one mentioned on the con-
trol cabinet and the power switch is in the Position 0.  

 

 

 

6.2. CONNECTION / DISCONNECTION OF THE MICROMOTOR  

Align the red dot of the micromotor socket to the red dot of the connector on the front face of the cabinet and clipper : 

 

 

 

 

 

 

 

 

 

 

Technical Specifications  

Dimensions  

 

Cabinet (w x d x h)

 

243 x 222 x 102 mm

 

Pedal (l x d x h)

 

280 x 170 x 70 mm

 

Micromotor (Ø x l)

 

Ø23 x 115 mm

 

Micromotor cable (l)

 

2 m 

Cabinet 

 

2.5 kg 

Weight  

Pedal 

 

1 kg 

Micromotor 

 

110 g 

Micromotor cable 

 

110 g 

Electrical  

Input power supply

 

from 100 to 240 V (50-60 Hz)

 

Power consumption

 

200 to 300W

 

Input fuse

 

Ø5 x 20 - T5A 250 V

 

Classification  

Cabinet

 

IP4X

 

Pedal

 

IPX6

 

Electrical classification  

Class

 

I according to IEC 60601-1

 

Type

 

Classification Bluetooth  

Group

 

1

 

Class

 

B

 

FCC ID

 

QOQBGM111

 

IC

 

5123A-BGM111

 

Micromotor  

Rotation speed

 

100 - 40000 rpm

 

Motor connection 

Model with light : TYPE 1-short 

Model without light : TYPE 4-short  

Peristaltic pump  

Maximum flow rate at 100% with Anthogyr line

 

120 ml/min

 

Pump pipe Øext

 

Ø7 mm 

Pump pipe Øint

 

Ø4 mm

 

CE classification  

Directive 93/42/CEE

 

Class IIa conforming to rule 8 for cabinet / pedal / micromotor ca-

ble  /  micromotor  class  I  conforming  to  rule  1  for  the  micromotor 

mount / clips

 

Display  

N.cm

 

Torque (Newton-centimetre)

 

rpm 

Speed (revolutions per minute)

 

6. CONNECTIONS - INSTALLATION  

Содержание Xpert Unit

Страница 1: ...bcdef ghij EN User manual ...

Страница 2: ...at you have placed in it In order to get complete and sustainable benefit from this product it is recommended that you read this user manual first and perform the recommended inspection and maintenance tasks This Anthogyr product is used exclusively with the accessories provided by Anthogyr micromotor pedal power supply cable The use of any other accessory is prohibited This equipment is compatibl...

Страница 3: ...f the manufacturers for the instrumentation used This is to limit the risk of injury and deterioration of instruments ATTENTION Once the servicing work is completed switch off the device Unplug it from the electrical mains in the event of prolonged non use WARNING a EN USER GUIDE Français English Deutsch Español Italiano Português Nederlands 中国的 عر 4 2 5 Storage conditions 4 2 6 Peristaltic pump 4...

Страница 4: ...e Position 0 6 2 CONNECTION DISCONNECTION OF THE MICROMOTOR Align the red dot of the micromotor socket to the red dot of the connector on the front face of the cabinet and clipper Technical Specifications Dimensions Cabinet w x d x h 243 x 222 x 102 mm Pedal l x d x h 280 x 170 x 70 mm Micromotor Ø x l Ø23 x 115 mm Micromotor cable l 2 m Cabinet 2 5 kg Weight Pedal 1 kg Micromotor 110 g Micromotor...

Страница 5: ...position I At startup all the pictograms on the screen light up and the device emits a beep The code Er02002 at startup means there is an anomaly with the date and time The date and time synchronise automatically with the mobile app when you are connected 7 2 WIRELESS PEDAL CONNECTION 7 2 1 At the 1st use At unit startup the pictogram display means that the pedal is on standby Press on the pedal l...

Страница 6: ...ngles of the MontBlanc Implant series 10400X XL 16400X XL With these contra angles the accuracy of torque is 10 for the range from 5 to 80 N cm 8 3 SETTING THE VALUES OF SPEED TORQUE FLOW RATE 8 3 1 Setting the rotation speed and the maximum torque 1 Pressing on the parameter to be changed speed or torque 2 The parameter starts flashing and the buttons appear 3 Adjust the value of the parameter wi...

Страница 7: ... flashes indicating the need to recharge soon see instructions below When charge level of the battery reaches 0 the device will beep every 30 seconds and it is not possible to use the motor Recharge according to the following instructions 8 7 RECHARGING THE PEDAL BATTERY Connect the charging adapter to the pedal through the connector located at the rear of the pedal Connect the adapter to the main...

Страница 8: ...ect for a period of 24 months from the date of invoice This warranty does not apply to wearing parts Any modification or addition to the product without the express consent of the company Anthogyr will nullify this warranty The warranty will lapse in case of non observance of the technical instructions provided with all our devices Anthogyr cannot be held responsible for damages and their conseque...

Страница 9: ... problem per sists do not use the motor pedal micro motor any longer Contact the Anthogyr After Sales Service for more infor mation Er 02 005 The touch bar encounters operating problems The micromotor pedal micromotor assembly can still be used safely with the pedal to complete the ongoing care Er 02 006 Er 02 007 Update the cabinet software using the mobile application Er 04 006 Overheating of th...

Страница 10: ...use any interference in nearby electronic equipment XPERT UNIT motor is suitable for use in all premises including domestic premises and those directly connected to the low voltage public electricity supply network feeding the buildings for domestic use Directives and declaration of the manufacturer Electromagnetic immunity The XPERT UNIT motor is intended to be used in the environment specified b...

Страница 11: ...gths are less than 3 V m Recommended separation distances between portable and mobile RF communication devices and the XPERT UNIT motor XPERT UNIT motor is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled Customer or user of the XPERT UNIT motor can contribute to prevent electromagnetic interferences by maintaining a minimum distance between the ...

Страница 12: ...redits Anthogyr All rights reserved Not contractual photos 16313000_NOT_2019 07 A Anthogyr SAS 2 237 Avenue André Lasquin 74700 Sallanches France Tél 33 0 4 50 58 02 37 Fax 33 0 4 50 93 78 60 www anthogyr fr K ...

Отзывы: