3-4 Spectrum Analyzer
Quick Start Guide
3-8
PN: 10580-00252 Rev. K
Site Master User Guide
Set the Amplitude
Press the
Amplitude
main menu key to display the Amplitude menu.
Set Amplitude Reference Level and Scale
1.
Press the
Reference Level
submenu key and use the arrow keys, rotary knob, or the
keypad to change the reference level. Press
Enter
to set the reference level value.
2.
Press the
Scale
submenu key and use the arrow keys, rotary knob, or the keypad to
enter the desired scale. Press
Enter
to set the scale value.
Set Amplitude Range and Scale
1.
Press the
Auto Atten
submenu key to set an optimal reference level based on the
measured signal.
2.
Press the
Scale
submenu key.
3.
Enter the desired scale units by using the keypad, the arrow keys, or the rotary knob.
Press
Enter
to set. The y-axis scale is automatically renumbered.
Power Offset Set Up for Compensating External Loss
To obtain accurate results, compensate for any external attenuation by using power offset.
In power offset mode, the compensation factor is in dB. (External attenuation can be created
by using an external cable or an external high power attenuator.)
Press the
RL Offset
submenu key and use the keypad, the arrow keys, or the rotary knob to
enter the desired offset value. When using the rotary knob, the value changes in increments
of 0.1 dB. Using the
Left/Right
arrow keys changes the value in 10% increments of the value
shown on the
Scale
submenu key. When using the
Up/Down
arrow keys, the value changes in
the increment shown on the
Scale
submenu key. When using the keypad, enter the new value
then press
Enter
or the
dB
submenu key to set the value. The power offset is displayed in the
instrument settings summary column on the left side of the measurement display.
Set the Span
1.
Press the
Span
main menu key or the
Freq
main menu key followed by the
Span
submenu key.
2.
To select full span, press the
Full Span
submenu key. Selecting full span overrides any
previously set Start and Stop frequencies.
3.
For a single frequency measurement, press the
Zero Span
submenu key.
Note
To quickly move the span value up or down, press the Span Up 1-2-5 or Span
Down 1-2-5 submenu keys These keys facilitate a zoom-in, zoom-out feature in a
1-2-5 sequence.
Содержание S331E
Страница 2: ......
Страница 36: ...2 9 Tilt Bail Stand Instrument Overview 2 22 PN 10580 00252 Rev K Site Master User Guide ...
Страница 48: ...3 4 Spectrum Analyzer Quick Start Guide 3 12 PN 10580 00252 Rev K Site Master User Guide ...
Страница 76: ...5 6 Updating the Site Master Firmware System Operations 5 12 PN 10580 00252 Rev K Site Master User Guide ...
Страница 82: ...7 1 Overview Bias Tee Option 10 7 2 PN 10580 00252 Rev K Site Master User Guide ...
Страница 98: ...9 3 Web Remote Control Interface Web Remote Control 9 12 PN 10580 00252 Rev K Site Master User Guide ...
Страница 100: ...A 1 Introduction Measurement Guides A 2 PN 10580 00252 Rev K Site Master User Guide ...
Страница 124: ...Index 4 PN 10580 00252 Rev K Site Master User Guide ...
Страница 125: ......