Quick Start Guide
3-3 Cable & Antenna Analyzer
Site Master User Guide
PN: 10580-00252 Rev. K
3-5
DTF Setup
1.
Press the
Measurements
main menu key and select
DTF Return Loss
or
DTF VSWR
.
2.
Press the
Freq/Dist
main menu key.
3.
Press the
Units
submenu key and select
m
to display distance in meters or
ft
to display
distance in feet.
4.
Press
DTF Aid
and use the touch screen or arrow keys to navigate through all the DTF
parameters.
a.
Set
Start Distance
and
Stop Distance
. Stop Distance needs to be smaller than
Dmax.
b.
Enter the
Start
and
Stop
frequencies.
c.
Press
Cable
, select the appropriate cable from the cable list and press
Enter
.
d.
Press
Continue
.
5.
Press
Shift
, then
Calibrate
(2) to calibrate the instrument. Refer to
for additional information.
6.
Press the
Marker
main menu key and set the appropriate markers.
7.
Press
Shift
and
Limit
(6) to enter and set the appropriate limit lines.
8.
Press
Shift
and
File
(7) to save the measurement. See the User Guide for details.
Figure 3-4.
DTF Aid
Содержание S331E
Страница 2: ......
Страница 36: ...2 9 Tilt Bail Stand Instrument Overview 2 22 PN 10580 00252 Rev K Site Master User Guide ...
Страница 48: ...3 4 Spectrum Analyzer Quick Start Guide 3 12 PN 10580 00252 Rev K Site Master User Guide ...
Страница 76: ...5 6 Updating the Site Master Firmware System Operations 5 12 PN 10580 00252 Rev K Site Master User Guide ...
Страница 82: ...7 1 Overview Bias Tee Option 10 7 2 PN 10580 00252 Rev K Site Master User Guide ...
Страница 98: ...9 3 Web Remote Control Interface Web Remote Control 9 12 PN 10580 00252 Rev K Site Master User Guide ...
Страница 100: ...A 1 Introduction Measurement Guides A 2 PN 10580 00252 Rev K Site Master User Guide ...
Страница 124: ...Index 4 PN 10580 00252 Rev K Site Master User Guide ...
Страница 125: ......