3-3 Cable & Antenna Analyzer
Quick Start Guide
3-6
PN: 10580-00252 Rev. K
Site Master User Guide
Calibrate with OSL Calibration
1.
Press the
Freq/Dist
main menu key and enter the appropriate frequency range
2.
Press
Shift
, followed by
Calibrate
(2).
3.
Select
Standard
or
FlexCal
.
4.
Press
Start Cal
and follow instructions on screen.
5.
Connect Open to
RF Out
and press the
Enter
key.
6.
Connect Short to
RF Out
and press the
Enter
key.
7.
Connect Load to
RF Out
and press the
Enter
key.
8.
Verify that the calibration has been properly performed by checking that the Cal Status
message is now displaying “
ON, Standard
” or “
ON, FlexCal
”.
Note
Refer to the Cable & Antenna Measurement Guide listed in
for
calibration details.
Figure 3-5.
Calibration Setup with OSL Cal
Power
Charge
+/-
.
0
3
Sweep
2
Calibrate
1
Preset
6
Limit
5
Trace
4
Measure
9
Mode
8
System
7
File
Shift
Esc
Enter
Menu
Site
Master
S362E
OPEN
LOAD
SHORT
CALIBRATION
RFOUT/REFLECTION
TEST PORT
TEST PORT CABLE (OPTIONAL)
Содержание S331E
Страница 2: ......
Страница 36: ...2 9 Tilt Bail Stand Instrument Overview 2 22 PN 10580 00252 Rev K Site Master User Guide ...
Страница 48: ...3 4 Spectrum Analyzer Quick Start Guide 3 12 PN 10580 00252 Rev K Site Master User Guide ...
Страница 76: ...5 6 Updating the Site Master Firmware System Operations 5 12 PN 10580 00252 Rev K Site Master User Guide ...
Страница 82: ...7 1 Overview Bias Tee Option 10 7 2 PN 10580 00252 Rev K Site Master User Guide ...
Страница 98: ...9 3 Web Remote Control Interface Web Remote Control 9 12 PN 10580 00252 Rev K Site Master User Guide ...
Страница 100: ...A 1 Introduction Measurement Guides A 2 PN 10580 00252 Rev K Site Master User Guide ...
Страница 124: ...Index 4 PN 10580 00252 Rev K Site Master User Guide ...
Страница 125: ......