MT9090A/MU909011A Network Master User’s Guide
7-7
Pulse Width
Check that the pulse width satisfies the specifications.
Connect the devices as shown in the following figure.
Figure 7-2: Pulse Width Test – Connection Diagram
Test Procedure:
1.
Start a Real-time measurement.
2.
Adjust the oscilloscope amplitude and time base
scale to display the waveform in the oscilloscope.
Adjust the optical variable attenuator to prevent the
waveform monitor from becoming saturated.
3.
Observe the oscilloscope waveform and measure the
peak level half amplitude as shown in the following
figure then check that measured results are with the
specifications.
Figure 7-3: Example – Peak Level at Half Amplitude measurement
MU909011
Series
Waveform
Monitor
Oscilloscope
Optical
Variable Attenuator
Optical
Fiber
Optical
Fiber
Peak Level
Pulse Width
100%
50%
Содержание Network Master MT9090A Series
Страница 32: ...xxxii MT9090A MU909011A Network Master User s Guide...
Страница 40: ...1 8 MT9090A MU909011A Network Master User s Guide...
Страница 68: ...2 28 MT9090A MU909011A Network Master User s Guide...
Страница 98: ...3 30 MT9090A MU909011A Network Master User s Guide...
Страница 142: ...4 44 MT9090A MU909011A Network Master User s Guide...
Страница 162: ...6 6 MT9090A MU909011A Network Master User s Guide...
Страница 180: ...7 18 MT9090A MU909011A Network Master User s Guide...
Страница 185: ...MT9090A MU909011A Network Master User s Guide 8 5...
Страница 186: ...8 6 MT9090A MU909011A Network Master User s Guide...
Страница 198: ...B 2 MT9090A MU909011A Network Master User s Guide portman 5 beta Other 5 Package Name License Remarks...