Section 4 Peformance Tests
4-14
(4)
Test procedure
Step
Operation procedure
1.
Set the signal generator to 101 MHz, –12 dBm.
2.
Execute Cal of the calibration receiver.
3.
Press the spectrum analyzer Preset key.
4.
Execute Cal.
5.
Set the MT8801C as follows:
Center Freq: 101 MHz Span: 3 MHz
Reference Level: 0 dBm Attenuator: 0 dB
RBW: 1 kHz
VBW: Minimum
6.
Press the Peak
→
REF key. The peak point of the spectrum waveform is set to the reference level.
7.
At the end of one sweep, press the Delta Marker key.
8.
Set the measured value of the calibration receiver to “Measure to Reference”.
9.
Read the delta marker level value (B) of the measured value (A) of the calibration receiver when the
output level value of the signal generator value is changed as shown in the table below.
10.
Repeat steps 5 through 9 in accordance with the table below.
LOG Linearity Test
Set value of output level
of signal generator (dB)
0
–10
–20
–30
–40
–50
–60
–70
–80
RBW=1 kHz
0
(Reference)
Error (B
−
A) (B)
RBW=30 kHz
0
(Reference)
RBW=1 MHz
0
(Reference)
Measurement
uncertainty
Effective
lower limit
Effective
upper limit
---
---
±
0.03 dB
±
0.03 dB
–
0.47 dB
+0.47 dB
±
0.04 dB
–
0.46 dB
+0.46 dB
–
0.96 dB
+0.96 dB
Содержание MT8801C
Страница 1: ...i MT8801C Radio Communication Analyzer Option 07 Spectrum Analyzer Operation Manual Panel Operation ...
Страница 3: ...1 1 Section 1 General 1 1 General 1 2 1 2 Manual Composition 1 3 1 3 Specifications 1 4 ...
Страница 8: ...Section 1 General 1 6 ...
Страница 10: ...Section 2 Panel Layout 2 2 ...
Страница 50: ...Section 3 Operation 3 40 Example of Waveform With Delay Time ...
Страница 82: ...Section 3 Operation 3 72 ...
Страница 116: ...Section 4 Peformance Tests 4 34 ...
Страница 117: ...App 1 Appendixes Appendix A Soft Key Menu A 1 Appendix B Keywords Index B 1 ...
Страница 118: ...Appendixes App 2 ...