Section 4 Peformance Tests
4-28
(4)
Test procedure
Step
Operation procedure
1.
Press the spectrum analyze Preset key.
2.
Execute Cal.
3.
Set the signal generator output to 100 MHz, 0 dBm.
4.
Set the spectrum analyzer as follows:
Center Freq: 100 MHz
Span: 15 kHz
Reference Level: 0 dBm
RBW: 3 kHz
Input terminal: Aux Input
5.
Press the Peak
→
CF key to set the peak frequency to the center frequency.
6.
Set the marker to Delta Marker.
7.
Set RBW and Span in accordance with the table below, and measure the level deviation for each RBW
as described in Step 8.
8.
Press the Peak Search key, and move the current marker to the spectrum waveform peak point. Read
the delta marker level.
Resolution Bandwidth Switching Deviation Test
+0.1 dB
RBW
300 Hz
1 kHz
3 kHz
10 kHz
30 kHz
100 kHz
300 kHz
1 MHz
Span
10 kHz
10 kHz
15 kHz
50 kHz
150 kHz
500 kHz
1.5 MHz
5 MHz
Rating
Spectrum analyzer setting
Delta marker
level
0.0 dB
(reference)
Measurement
uncertainty
±
0.01 dB
Effective
lower limit
Effective
upper limit
–0.09 dB
+0.09 dB
Содержание MT8801C
Страница 1: ...i MT8801C Radio Communication Analyzer Option 07 Spectrum Analyzer Operation Manual Panel Operation ...
Страница 3: ...1 1 Section 1 General 1 1 General 1 2 1 2 Manual Composition 1 3 1 3 Specifications 1 4 ...
Страница 8: ...Section 1 General 1 6 ...
Страница 10: ...Section 2 Panel Layout 2 2 ...
Страница 50: ...Section 3 Operation 3 40 Example of Waveform With Delay Time ...
Страница 82: ...Section 3 Operation 3 72 ...
Страница 116: ...Section 4 Peformance Tests 4 34 ...
Страница 117: ...App 1 Appendixes Appendix A Soft Key Menu A 1 Appendix B Keywords Index B 1 ...