Section 4 Peformance Tests
4-12
(4)
Test procedure
Step
Operation procedure
1.
Set the signal generator to 1 GHz, +10 dBm.
2.
Press the spectrum analyzer Preset key.
3.
Execute Cal.
4.
Set the MT8801C as follows:
Center Freq: 1 GHz
Span: 40 kHz (table below)
Reference Level: +10 dBm
Detection: Sample
RBW: 1 kHz (table below)
VBW: 10 Hz (table below)
Zone Width: Spot
5.
Press the Peak Search key. The current marker is set to the peak point.
6.
Press the Measure key, and select C/N Ratio Measure.
7.
Press the Meas On key, and execute C/N measurement.
8.
Press the Marker key. Then, turn the rotary knob to move the marker to the right so that the Zone
Center display frequency becomes the frequency shown in the table below.
9.
Check that the sideband noise value does not exceed the rated value.
10.
Repeat steps 4 through 9 in accordance with the table shown below.
Sideband Noise Level Test
RBW
VBW
Zone Center
(offset)
40 kHz
400 kHz
1 kHz
10 kHz
10 Hz
100 Hz
10 kHz
100 kHz
≤
–95.82 dBc
≤
–116.11 dBc
±
0.82
±
1.11
Frequency
span
Measured
value
Effective
upper limit
Sideband noise
Measurement
uncertainty
Содержание MT8801C
Страница 1: ...i MT8801C Radio Communication Analyzer Option 07 Spectrum Analyzer Operation Manual Panel Operation ...
Страница 3: ...1 1 Section 1 General 1 1 General 1 2 1 2 Manual Composition 1 3 1 3 Specifications 1 4 ...
Страница 8: ...Section 1 General 1 6 ...
Страница 10: ...Section 2 Panel Layout 2 2 ...
Страница 50: ...Section 3 Operation 3 40 Example of Waveform With Delay Time ...
Страница 82: ...Section 3 Operation 3 72 ...
Страница 116: ...Section 4 Peformance Tests 4 34 ...
Страница 117: ...App 1 Appendixes Appendix A Soft Key Menu A 1 Appendix B Keywords Index B 1 ...
Страница 118: ...Appendixes App 2 ...