4-3
4.2
Instruments Required for Performance Test
The instruments required for performance tests are shown below.
Instruments Required for Performance Test
Test item
Center frequency display
accuracy
Frequency span display
accuracy
Resolution bandwidth and
selectivity
Sideband noise level
Screen amplitude display
linearity
Frequency response
Reference level accuracy
Average noise level
Second harmonic distortion
Resolution bandwidth
switching deviation
Test equipment name
Signal generator
Signal generator
Signal generator
Calibration receiver
Signal generator
Power meter
Power sensor
Signal generator
Calibration receiver
Power meter
Power sensor
No-reflection
terminator
Recommended model
MG3633A
MG3633A
MG3633A
ML2530A
HP8665B
ML4803A
MA4601A
6769A
ML2530A
ML4803A
MA4601A
MP752A
Reference
4.3.2
Signal generator
MG3633A
4.3.3
Signal generator
MG3633A
4.3.4
4.3.5
4.3.6
4.3.7
4.3.8
4.3.9
4.3.11
Signal generator
MG3633A
4.3.12
Low-pass filter
Fundamental frequency (Device
with a attenuation of at least 70
dB at double the frequencies 10
MHz and 1 GHz)
Residual response
50
Ω
terminator
MP752A
4.3.10
4.2 Instruments Required for Performance Test
Содержание MT8801C
Страница 1: ...i MT8801C Radio Communication Analyzer Option 07 Spectrum Analyzer Operation Manual Panel Operation ...
Страница 3: ...1 1 Section 1 General 1 1 General 1 2 1 2 Manual Composition 1 3 1 3 Specifications 1 4 ...
Страница 8: ...Section 1 General 1 6 ...
Страница 10: ...Section 2 Panel Layout 2 2 ...
Страница 50: ...Section 3 Operation 3 40 Example of Waveform With Delay Time ...
Страница 82: ...Section 3 Operation 3 72 ...
Страница 116: ...Section 4 Peformance Tests 4 34 ...
Страница 117: ...App 1 Appendixes Appendix A Soft Key Menu A 1 Appendix B Keywords Index B 1 ...