Anest Iwata ISP-50 Скачать руководство пользователя страница 20

 

-18- 

Check cooling air 

    Check that cooling fan is turning and cooling air is flowing in that direction as indicated. 

 
 
 
 
 
 
 
 
 
 
 
 

 

3.3 Connection to vacuum system (chamber) 

Inlet of ISP-50 is NW25 and outlet is NW16. 

 

  

 

CAUTION

 

Danger of exhaust disruption 

 

When connecting exhaust piping to vacuum pump and when combining piping 
with another vacuum pump, pay attention to piping size and length so that it does 
not cause exhaust resistance. 

Exhaust resistance can disrupt air flow, resulting in failure and overcurrent.

 

 

Pay attention to 

exhaust resistance 

 

Important     

Use isolation valve between vacuum system and inlet. 

Isolation valve is necessary to prevent the drawback of debris attached to the inside of vacuum pump into the 
vacuum chamber during start-up and shut-down. (We recommend the use of leak valve also). We recommend 
the use of an 

automatic valve as the isolation valve which closes during power failure in order to prevent 

the drawback of debris inside pump into the vacuum chamber during power failure.

 

 

Use the clean connecting pipe between vacuum chamber and vacuum pump. 

We recommend the use of a flexible tube between inlet of vacuum pump and vacuum chamber so that vibration 
of pump does not transmit to vacuum chamber. 
 

When connecting exhaust piping to outlet of vacuum pump, refer to the following size and length. 

It is recommended in the case of ISP-50, max. 

30m

 direct pipe length for exhaust pipe size 

NW16 (inner dia.16) 

But if pipe length becomes longer, use a larger size exhaust pipe. 

 

Make sure that exhaust piping is not clogged during pump operation. 

Make sure that pressure at outlet does not exceed atmospheric pressure at any conditions. 

 
 
 
 
 
 
 
 
 

Chamber 

Isolation valve 

(Automatic valve) 

Leak 

valve 

Vacuum pump 

Outlet duct 

Inlet 

Outlet 

ISP-50 

Cooling fan 

Inlet of cooling fan 

Outlet of cooling air 

Содержание ISP-50

Страница 1: ...urchase date When you began operation This instruction manual includes very important warnings cautions and operating procedure in order to operate this pump safely and efficiently Be sure to read thi...

Страница 2: ......

Страница 3: ...d marks have the following meanings Examples of marks WARNING Indicates a potentially hazardous situation which if not avoided may result in serious injury or loss of life CAUTION Indicates a potentia...

Страница 4: ...jury due to turning objects Install overcurrent protective device Danger of accident fire and failure Be sure to install protective device to protect circuitry We recommend an overcurrent protective d...

Страница 5: ...n cause bodily injury from entanglement with turning section or foreign matter dispersal Never alter Danger of electric shock and entanglement Do not remove or alter safeguards or insulation parts If...

Страница 6: ...Danger of problem recurrence and failure If protective device or thermal protector activates be sure to turn off electric source and inspect causes to solve the problem Do not operate until problem i...

Страница 7: ...nger of remaining moisture When evacuating moisture be sure to open air flush port air flush operation If you evacuate vapor while air flush port is closed condensed moisture will remain inside vacuum...

Страница 8: ...s keep warning stickers clean and legible If they become dirty or detached replace them with new ones If you need replacement stickers contact the dealer who sold the vacuum pump to you Motor frame Re...

Страница 9: ...er 18 4 Operation 19 4 1 Standard operation 21 4 1 1 Start up 21 4 1 2 Shut down 21 4 2 Air flush operation 22 4 2 1 Preparation 22 4 2 2 Start up and shut down 23 4 2 3 When returning to standard ope...

Страница 10: ...side up before opening Check that the model of the product is the same as the one you ordered How to read model name I S P 5 0 S V 1 ISP model Inlet direction V vertical Motor S Single phase Displace...

Страница 11: ...itten by official languages except English must be sent to a customer along with the delivery of a product Air muffler for air flush ISP 50 is attached to outlet flange Please prepare electric wires c...

Страница 12: ...m pump If not it can cause damage failure or bodily injury from falling vacuum pump or by being caught between vacuum pump and other material Be careful about lifting This product does not come with a...

Страница 13: ...11 2 Name and structure of each section ISP 50 Motor power box Inlet NW25 Turning direction plate Motor Outlet NW16 Air Flush Port Rc 1 8 Cooling fan Receptacle Hour meter...

Страница 14: ...f 5 40 Operating at a temperature range other than that designated can cause accident failure or bodily injury such as burns due to overheating Use at designated temperature Danger of overheating Inst...

Страница 15: ...ectrician to perform electric wiring If not short circuit or electric shock can cause fire or bodily injury Ask qualified electrician Danger of electric shock and entanglement Be sure to turn off elec...

Страница 16: ...n cause short circuit fire or bodily injury from electric shock due to looseness or disconnection Be careful about insertion Danger of electric shock Connect grounding wire to electric source If not i...

Страница 17: ...ge Danger of problem recurrence and failure If motor protective device or thermal protector activates be sure to turn off electric source and inspect causes to solve the problem Do not operate until p...

Страница 18: ...epare electric wire with a connector corresponding to the receptacle table 2 Protective device must be installed between this cord and electric source Insert a connector of electric wire into receptac...

Страница 19: ...ure Prevent foreign matter from entering Danger of overheating Check that cooling fan is turning and cooling air is flowing If not it can cause accident failure or bodily injury such as burns due to o...

Страница 20: ...recommend the use of an automatic valve as the isolation valve which closes during power failure in order to prevent the drawback of debris inside pump into the vacuum chamber during power failure Use...

Страница 21: ...inger or foreign matter into air hole of fan cover air hole of motor or clearance between FS 1 and FS 2 cooling fins If done it can cause bodily injury from entanglement with turning section or foreig...

Страница 22: ...ndensed water will remain inside vacuum pump and cause failure Operate while opening air flush port Danger of insufficient vapor exhaust After evacuating vapor do air flush operation for at least one...

Страница 23: ...cause failure When doing air flush operation Noise level and ultimate pressure will increase by 7 8dB about 5Pa Install pump in an area which is not exposed to debris such as iron powder stone powder...

Страница 24: ...ng vapor You must shorten maintenance interval 5 2 page 25 when evacuating vapor since vapor temperature disposal volume disposal frequency and substances in vapor have an influence on pump operation...

Страница 25: ...sure to close isolation valve between vacuum pump and vacuum chamber during pump shut down If vacuum pump stops during air flush operation atmospheric air is drawn back from air flush port to inside...

Страница 26: ...uct maintenance and inspection after you turn off electric source If not it can cause bodily injury from electric shock or turning object Turn off electric source Danger of accident failure and shorte...

Страница 27: ...very8 000 hours Every2 yearor every16 000hours Every400times of vapor pumping Bearing kit grease Tip seal set Seal set O ring set Exhaust valve set Air flush kit Pin crank kit Vacuum pump itself Insid...

Страница 28: ...t and repair Improper exhaust piping Check exhaust piping diameter and length Air outlet is clogged Remove blank flange from exhaust outlet Motor stops soon Motor protection gear activates Air outlet...

Страница 29: ...27 7 Disposal When a vacuum pump is disposed please comply with local law such as the Waste Disposal Law...

Страница 30: ...et connection NW16 Direction of inlet Vertical Dimensions mm L W H 317 155 227 Mass kg 12 Cooling system Air cooled O t h e r s With hour counter and air flush Note 1 Pumping speed and ultimate pressu...

Страница 31: ...29 8 2 Dimensions 8 2 1 ISP 50 SV1...

Страница 32: ...30 8 3 Performance data...

Страница 33: ...memo...

Страница 34: ...rer 3176 Shinyoshida cho Kohoku Ku Yokohama 223 8501 Japan Tel 81 0 45 591 1112 Fax 81 0 45 591 1539 http www anest iwata co jp European agent Corso Vigevano 56 10155 Torino ITALY Tel 39 1 1248 0868 F...