INTRODUCTION
Neo
SECTION 1
Page 21
1.5.4 - Cooling pipe connectors
Two connectors are fitted to the camera in order to allow water cooling pipes to be connected and these can be
connected to a water cooler or recirculator to improve cooling:
Figure 4: Cooling pipe connectors
Please refer to the mechanical drawings in
Section 1.5.1
for details of connector and hose type compatibility
and to
Section 5.4
for connection & disconnection information.
Some mains supply water is heavily mineralised, (i.e. “Hard”) which could cause deposits in the water circuit
inside the camera. This can reduce the flow-rate and effect the cooling achieved, therefore to prevent deposits
forming, it is recommended that de-ionized water (without additives) is used as the coolant.
The specified cooling performance of the camera can be achieved with coolant flow rates of 2 litres per minute,
the maximum recommended pressure of coolant circulating through the camera head is 2 bar.
NOTE: Always ensure that the temperature of liquid coolant circulated through the camera head is
above the dew point of the camera ambient. Use of coolant at or below the dew point will result in
permanent damage to the camera head, due to formation of condensation on internal components.