Operation Guide 

Template version:1

 

9/14 

7.  TROUBLESHOOTING (1) 

 

PROBLEM 

POSSIBLE CAUSE 

SOLUTION 

LCD Display shows   

battery symbol 

Low Battery

 

Change all the batteries

 

LCD Display shows 

 

“EE” 

Arm or blood pressure 

monitor was moved 

during testing 

Re-test taking care to not 

move your arm or the 

blood pressure monitor 

The cuff does not inflate 

properly or pressure falls 

quickly during testing 

Make certain the rubber 

tube is fully inserted into 

the blood pressure 

monitor 

Irregular heartbeat 

(arrhythmia) 

It is inappropriate for 

people with serious 

arrhythmia to use this 

blood pressure monitor. 

 

8.  TROUBLESHOOTING (2) 

 

PROBLEM 

POSSIBLE CAUSE 

SOLUTION 

LCD Display shows 

 

“EE” 

The cuff was not properly 

applied or the rubber 

tube was bent or 

pressed. 

Review the cuff applying 

and testing sections of 

the instructions and 

re-test. 

LCD Display shows 

abnormal result 

The cuff position was not 

correct or it was not 

properly tightened 

Apply the cuff correctly 

and try again 

Body posture was not 

correct during testing 

Review the body posture 

and testing sections of 

the instructions and 

re-test. 

Speaking, arm or body 

movement, angry, 

excited or nervous during 

testing 

Re-test when calm and 

without speaking or 

moving during the test 

No response when you 

press button or load 

battery

 

Incorrect operation, or 

strong electromagnetic 

interference 

Take out batteries for five 

minutes, and then 

reinstall all batteries. 

Содержание KD-322

Страница 1: ...RS 2 INTENDED USE 3 CONTRAINDICATION 3 PRODUCT DESCRIPTION 3 SPECIFICATIONS 4 NOTICE 4 SETUP AND OPERATING PROCEDURES 6 1 BATTERY LOADING 6 2 CONNECTING THE CUFF TO THE MONITOR 6 3 APPLYING THE CUFF 6...

Страница 2: ...The highest value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most peop...

Страница 3: ...egrated pressure sensor blood pressure and pulse rate can be measured automatically and non invasively The LCD display will show blood pressure and pulse rate The Electronic Blood Pressure Monitor cor...

Страница 4: ...on 90 RH 13 Environmental temperature for storage and transport 20 55 4 131 14 Environmental humidity for storage and transport 90 RH 15 Environmental pressure 80kPa 105kPa 16 Battery life Approx 540...

Страница 5: ...ted that you consult with your physician for accurate assessment 13 Please do not use the cuff other than supplied by the manufacturer otherwise it may bring biocompatible hazard and might result in m...

Страница 6: ...e 2 CONNECTING THE CUFF TO THE MONITOR Insert the cuff tubing connector into the socket in the left side of the monitor Make certain that the connector is completely inserted to avoid air leakage duri...

Страница 7: ...rtably Measurement a Be seated with your feet flat on the floor and don t cross your legs b Place palm upside in front of you on a flat surface such as a desk or table c The middle of the cuff should...

Страница 8: ...f the monitor manually h During measurement you can press the exhaust valve which is located in the front of the bulb to release the air Note Please consult a health care professional for interpretati...

Страница 9: ...itor 8 TROUBLESHOOTING 2 PROBLEM POSSIBLE CAUSE SOLUTION LCD Display shows EE The cuff was not properly applied or the rubber tube was bent or pressed Review the cuff applying and testing sections of...

Страница 10: ...iately qualified technical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a mini...

Страница 11: ...use in the electromagnetic environment specified below The customer or the user of the KD 322 should assure that it is used in such an environment Emissions test Compliance Electromagnetic environmen...

Страница 12: ...equency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Table 3 For ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING Gu...

Страница 13: ...radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic enviro...

Страница 14: ...12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be de...

Отзывы: