Andon KD-322 Скачать руководство пользователя страница 6

  DN:KD-322-SMSY01 V4.0 

Operation Guide 

Template version:1

 

6/14 

 

SETUP AND OPERATING PROCEDURES 

 

1.  BATTERY LOADING 

 

a.  Open battery cover at the back of the monitor. 

b. 

Load four “AA” size batteries. Please pay attention to polarity. 

c.  Close the battery cover. 

When LCD shows battery symbol

, replace all batteries with new ones.   

Rechargeable batteries are not suitable for this monitor. 

Remove  the  batteries  if  the  monitor  will  not  be  used  for  a  month  or  more  to  avoid 

relevant damage of battery leakage. 

Avoid the battery fluid to get in your eyes. If it should get in your eyes, immediately 

rinse with plenty of clean water and contact a physician. 

The monitor, the batteries and the cuff, must be disposed of according to local 

regulations at the end of their usage. 

 

2.  CONNECTING THE CUFF TO THE MONITOR 

 

Insert the cuff tubing connector into the socket in the left side of the monitor.    Make 

certain that the connector is completely inserted to avoid air leakage during blood pressure 

measurements.

 

 

 

 

 

 

Avoid compression or restriction of the connection tubing during measurement

which may cause inflation error, or harmful injury due to continuous cuff pressure

.

 

 

3. 

  APPLYING THE CUFF

 

 

a.  Pulling  the  cuff  end    through  the  medal  loop  (the  cuff  is 

packaged  like  this  already),  turn  it  outward  (away  from  your 

body) and tighten it and close the Velcro fastener. 

b.  Place the cuff around a bare arm 1-2cm above the elbow joint.   

c.  While seated, place palm upside in front of you on a flat surface 

such as a desk or table. Position the air tube  in  the  middle of 

your arm in line with your middle finger. 

d.  The  cuff  should  fit  comfortably,  yet  snugly  around  your  arm. 

You should be able to insert one finger between your arm and the cuff.   

Note:

   

Содержание KD-322

Страница 1: ...RS 2 INTENDED USE 3 CONTRAINDICATION 3 PRODUCT DESCRIPTION 3 SPECIFICATIONS 4 NOTICE 4 SETUP AND OPERATING PROCEDURES 6 1 BATTERY LOADING 6 2 CONNECTING THE CUFF TO THE MONITOR 6 3 APPLYING THE CUFF 6...

Страница 2: ...The highest value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most peop...

Страница 3: ...egrated pressure sensor blood pressure and pulse rate can be measured automatically and non invasively The LCD display will show blood pressure and pulse rate The Electronic Blood Pressure Monitor cor...

Страница 4: ...on 90 RH 13 Environmental temperature for storage and transport 20 55 4 131 14 Environmental humidity for storage and transport 90 RH 15 Environmental pressure 80kPa 105kPa 16 Battery life Approx 540...

Страница 5: ...ted that you consult with your physician for accurate assessment 13 Please do not use the cuff other than supplied by the manufacturer otherwise it may bring biocompatible hazard and might result in m...

Страница 6: ...e 2 CONNECTING THE CUFF TO THE MONITOR Insert the cuff tubing connector into the socket in the left side of the monitor Make certain that the connector is completely inserted to avoid air leakage duri...

Страница 7: ...rtably Measurement a Be seated with your feet flat on the floor and don t cross your legs b Place palm upside in front of you on a flat surface such as a desk or table c The middle of the cuff should...

Страница 8: ...f the monitor manually h During measurement you can press the exhaust valve which is located in the front of the bulb to release the air Note Please consult a health care professional for interpretati...

Страница 9: ...itor 8 TROUBLESHOOTING 2 PROBLEM POSSIBLE CAUSE SOLUTION LCD Display shows EE The cuff was not properly applied or the rubber tube was bent or pressed Review the cuff applying and testing sections of...

Страница 10: ...iately qualified technical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a mini...

Страница 11: ...use in the electromagnetic environment specified below The customer or the user of the KD 322 should assure that it is used in such an environment Emissions test Compliance Electromagnetic environmen...

Страница 12: ...equency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Table 3 For ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING Gu...

Страница 13: ...radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic enviro...

Страница 14: ...12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be de...

Отзывы: