INTENDED USE 

 

Semi-Automatic Blood Pressure Monitor is for use by medical professionals or at home and is 

a non-invasive blood pressure measurement system intended to measure the diastolic and 

systolic blood pressures and pulse rate of an adult individual by using a non-invasive 

technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference 

is limited to 22cm-48cm (approx. 8 21/32

” -18 29/32”). 

 

CONTRAINDICATION 

 

It is inappropriate for people with serious arrhythmia to use this electronic blood pressure 

monitor. 

 

PRODUCT DESCRIPTION 

 

Based on Oscillometric methodology and silicon integrated pressure sensor, blood pressure 

and pulse rate can be measured automatically and non-invasively. The LCD display will show 

blood pressure and pulse rate. The Electronic Blood Pressure Monitor corresponds to the 

below standards: IEC 60601-1:2005/EN 60601-1:2006/AC:2010 (Medical electrical equipment 

-- Part 1: General requirements for basic safety and essential performance), 

IEC60601-1-2:2007/EN 60601-1-2:2007 /AC:2010 (Medical electrical equipment -- Part 1-2: 

General requirements for basic safety and essential performance - Collateral standard: 

Electromagnetic compatibility - Requirements and tests), EN 1060-1: 1995 + A1: 2002 + A2: 

2009 (Non-invasive sphygmomanometers - Part 1: General requirements), EN 1060-3: 1997 + 

A1: 2005 + A2: 2009 (Non-invasive sphygmomanometers - Part 3: Supplementary 

requirements for electro-mechanical blood pressure measuring systems), ANSI/AAMI 

SP-10:2002+A1:2003+A2:2006. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание KD-322

Страница 1: ...RS 2 INTENDED USE 3 CONTRAINDICATION 3 PRODUCT DESCRIPTION 3 SPECIFICATIONS 4 NOTICE 4 SETUP AND OPERATING PROCEDURES 6 1 BATTERY LOADING 6 2 CONNECTING THE CUFF TO THE MONITOR 6 3 APPLYING THE CUFF 6...

Страница 2: ...The highest value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most peop...

Страница 3: ...egrated pressure sensor blood pressure and pulse rate can be measured automatically and non invasively The LCD display will show blood pressure and pulse rate The Electronic Blood Pressure Monitor cor...

Страница 4: ...on 90 RH 13 Environmental temperature for storage and transport 20 55 4 131 14 Environmental humidity for storage and transport 90 RH 15 Environmental pressure 80kPa 105kPa 16 Battery life Approx 540...

Страница 5: ...ted that you consult with your physician for accurate assessment 13 Please do not use the cuff other than supplied by the manufacturer otherwise it may bring biocompatible hazard and might result in m...

Страница 6: ...e 2 CONNECTING THE CUFF TO THE MONITOR Insert the cuff tubing connector into the socket in the left side of the monitor Make certain that the connector is completely inserted to avoid air leakage duri...

Страница 7: ...rtably Measurement a Be seated with your feet flat on the floor and don t cross your legs b Place palm upside in front of you on a flat surface such as a desk or table c The middle of the cuff should...

Страница 8: ...f the monitor manually h During measurement you can press the exhaust valve which is located in the front of the bulb to release the air Note Please consult a health care professional for interpretati...

Страница 9: ...itor 8 TROUBLESHOOTING 2 PROBLEM POSSIBLE CAUSE SOLUTION LCD Display shows EE The cuff was not properly applied or the rubber tube was bent or pressed Review the cuff applying and testing sections of...

Страница 10: ...iately qualified technical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a mini...

Страница 11: ...use in the electromagnetic environment specified below The customer or the user of the KD 322 should assure that it is used in such an environment Emissions test Compliance Electromagnetic environmen...

Страница 12: ...equency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Table 3 For ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING Gu...

Страница 13: ...radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic enviro...

Страница 14: ...12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be de...

Отзывы: