Intended use 

 

  AG-605A Blood glucose meter 

 

Intended use in home by lay person or by professional to monitor blood glucose 

 

Intended use for IN-Vitro Diagnostic. 

 

Intended use for monitoring blood glucose from capillary vessel whole blood(fingertip) 

 

Used with AGS-1000 test strip only 

Principle 

Glucose in blood sample mixes with chemicals on the test strip producing a small 

electrical current.

 

Product Brief Instruction 

AG-605A blood glucose Monitoring System is designed to provide an easy, accurate 
method  for  determination  of  your  blood  glucose  level  within  5  seconds.  350  test 
records with respective date and time help to review the blood glucose trend easily. 

Specification 
Meter Specification 

1.

 

Model: AG-605A 

2.

 

Machine size: 112mm(L)×50mm(W)×22mm(H) 

3.

 

Weight: 60g (exclude batteries) 

4.

 

Measuring method: Amperometric technology using glucose oxidase   

5.

 

Measuring range: 20 mg/dL ~600 mg/dL(1.1 mmol/L ~33.3mmol/L) 

6.

 

Measurement unit: mg/dL or mmol/L 

7.

 

Memory volume: 350times with time and date displaying 

8.

 

Display: LCD display 

9.

 

Power source: DC 3V (AAA×2 batteries) 

10.

 

Battery Life : Approx. 1000 normal tests 

11.

 

Storage Condition: -20

℃~

55

, Humidity

80%RH 

12.

 

Blood Source: Fresh capillary whole blood 

13.

 

Blood Volume: Minimum 0.7 micro liters. 

14.

 

Use life

five years   

Note

:  If  you  want  to  know  the  leave  factory  date  and  number  of  the  meter  please  check  the 

package. 

System Operation conditions:

   

Operation conditions: 10

℃~

40

, Humidity:25%~80%RH 

Warning: 

AG-605A Blood Glucose Meter should be used with 

AGS-1000 

test strip. 

 

List of AG-605A Blood Glucose Monitoring System 

 

Blood glucose meter 

 

instruction for use 

 

Sampler 

 

Lancets 

 

Test strip 

 

Quick Reference Guide 

 

Test strip indication for use 

Содержание AG-605A

Страница 1: ...m if it is not working properly or if it has suffered any damage Do not let the Blood Glucose Monitoring System come into contact with surfaces which are too hot to touch Do not block test ports nor p...

Страница 2: ...mm L 50mm W 22mm H 3 Weight 60g exclude batteries 4 Measuring method Amperometric technology using glucose oxidase 5 Measuring range 20 mg dL 600 mg dL 1 1 mmol L 33 3mmol L 6 Measurement unit mg dL o...

Страница 3: ...test strip Avoid making the liquid into the test strip and the button Avoid getting water or any other solution inside the meter Blood Sample The fresh capillary vessel whole blood finger tip blood or...

Страница 4: ...ize batteries and confirms well positive and negative extremely 3 Slide battery cover back into place Note Discard the used battery according to the local law Test Strip The Test Strips are designed t...

Страница 5: ...the date setting Short press and release the MEM button until the correct date appears 5 With the correct date on the display short press and release the S button to bring up the time setting The hou...

Страница 6: ...ired Testing Your Blood Glucose Level Before testing you will need the following items Glucose AG 605A Meter AGS 1000 test strip Adjustable Automatic Lancing Device Sterile Lancet CAUTION To reduce th...

Страница 7: ...Replace the end cap back on the body of the lancing device 5 Pull back on the cocking mechanism until it comes to a stop 6 Place the lancing device firmly against the sample site and press the release...

Страница 8: ...f the same result occurs consult your healthcare professional immediately Step 10 If a LO message appears on your display your Glucose Meter has detected that your blood glucose level is lower than 20...

Страница 9: ...t strip Control solution If long press the S button for about two seconds in step5 the CTL will appear on the screen which means that the test is under the CTL condition and the result will not be sto...

Страница 10: ...aning the andon Blood Glucose Monitoring System gently wipe the exterior surface using a damp soft cloth AVOID USING ANY ORGANIC SOLVENT for cleaning Discard the used lancet properly in accordance wit...

Страница 11: ...countdown The meter is calculating the result At the end of the count down the meter will display the test result No action required Blood glucose level is lower than 20mg dl 1 1mmol L The message in...

Страница 12: ...12 Display the result Do nothing Demonstrates the memory the serial number Do nothing Memory result demonstration Do nothing No memory in meter Do nothing Result display is end Do nothing...

Страница 13: ...libration Please contact us Use a used test strip Please use a new strip and re test Meter internal memory unit error Reinsert the CODE bottle please contact us if it appear for three time Environment...

Страница 14: ...e number printed on the Test Strip vial Incorrect Code numbers has been inputted into the Meter Insert the new CODE card to match the Code number on the Test Strip vial label Test results are inconsis...

Страница 15: ...15 15minutes before performing a new test The Meter countdown did not start Test Strip has not been inserted correctly Use a new Test Strip and redo the test...

Страница 16: ...erature limitation Expired Date Anti garbage Lotus Global Co Ltd 47 Spenlow House Bermondsey London SE16 4SJ Andon Health Co Ltd Register Address No 3 JinPing Street Ya An Road Nankai District Tianjin...

Отзывы: