![AND Weighing Indicator AD-4401 Скачать руководство пользователя страница 63](http://html1.mh-extra.com/html/and/weighing-indicator-ad-4401/weighing-indicator-ad-4401_instruction-manual_2941326063.webp)
61
INTERFACE SPECIFICATION
Output
Open collector (Common is connected to the NPN transistor emitter sides)
Output data: 6-digit BCD code, overflow, polarity, motion, strobe
Input
Dry contact (Pulled up to +5 V by a resistor)
Input data: Hold
• In case of negative logic, each terminal is turned on at “1”. “ON” refers to the state that there is power continuity
between the terminal and GND.
Example: The B14 terminal is turned on in NET.
•
In case of overflow the digits, 1x10
0
to 1x10
5,
of the weighing data become “1”.
•
Hold-input is fixed at negative logic. All outputs including strobe are held when HOLD input is turned on.
Pin #
Pin #
A1
1 x 10
0
B1
2 x 10
0
A2
4 x 10
0
B2
8 x 10
0
A3
1 x 10
1
B3
2 x 10
1
A4
4 x 10
1
B4
8 x 10
1
A5
1 x 10
2
B5
2 x 10
2
A6
4 x 10
2
B6
8 x 10
2
A7
1 x 10
3
B7
2 x 10
3
A8
4 x 10
3
B8
8 x 10
3
A9
1 x 10
4
B9
2 x 10
4
A10
4 x 10
4
B10
8 x 10
4
A11
1 x 10
5
B11
2 x 10
5
A12
4 x 10
5
B12
8 x 10
5
A13
OVER
B13 Positive
A14
Motion Detect
B14 Net
A15
Decimal Point 10
1
B15
Decimal Point 10
2
A16
Decimal Point 10
3
B16
Decimal Point 10
4
A17
* Unit 1
B17
* Unit 2
A18 Strobe B18
Hold-Input
A19 Common B19
Common
A20 FG B20
FG
*Unit 1, Unit 2
Unit
Unit 1
Unit 2
None 0
0
kg
0
0
lb
1
0
t
0
1
g
1
1
Содержание Weighing Indicator AD-4401
Страница 1: ...AD 4401 WEIGHING INDICATOR 1386 2A IE...
Страница 42: ...40...
Страница 87: ...85 APPENDIX DIMENSIONS AD 4401 OP 05...
Страница 88: ...86...
Страница 89: ...87...
Страница 90: ...88...